3.6 Connect the tubes
Arrow symbols identify the inlet and outlet ports for the syringe and the pump
connections. The “up” arrow is the outlet port. The “down” arrow is the inlet port. Turn the
tube connectors on the inlet and outlet ports of the syringe and pump until they click.
The suction tube that drains into the waste bottle must be installed on the bottom of the
beaker for good drainage.
N O T I C E
The diffusion tip and tube holder on the outlet tube, are pre-installed at the optimal positions. Do
not change the position of the diffusion tip or the tube holder.
3.7 Install the sensor
3.7.1 Connect the sensor
Use a conical adapter to hold the sensor tightly in the sensor holder. Make sure the
conical adapter is installed correctly. This makes sure of a secure water-tight fitting in the
sensor holder and the tip of the sensor will be in the correct position in the beaker.
Connect the sensor to an available sensor port on the rear of the instrument. After the
sensor is connected, make sure that the sensor icon shows in the banner at the top of the
display.
Refer to the illustrated steps that follow.
Installation
14