Section 6
Startup
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to the
current safety data sheets (MSDS/SDS) for safety protocols.
C A U T I O N
Personal injury hazard. Never use the instrument without the syringe cover installed.
6.1 Configure the instrument
1.
From the home screen, select
Settings
on page 22.
2.
Select an option, then push
Select
.
Option
Description
Applications
Changes (refer to
Change the application settings
on page 35), makes
copies, exports and removes application data. Make sure that the
duplication function does not make more than five applications.
Operators
Adds, changes and removes operators.
Date + Time
Sets the instrument date and time.
Brightness
Sets the brightness of the display.
Sounds
Sets the sound options.
Language
Sets the language.
Network
Give a name to the instrument. If a printer is connected, this name is
printed on the hard-copy output. Restart the instrument if the name is
changed.
Info
Shows information about the instrument and the attached hardware.
Restore
Defaults
Sets the instrument to the default configuration. The restored settings are:
•
The language is set to English
•
All password security is removed
•
The security password is set to 0000
•
No sound on: Key press, Information and Analysis results OK
•
Sound on: Instrument ready, Instrument switch off, Errors and
Analysis results not OK
•
The application parameters view is set to basic mode
•
Sets the syringe to empty when the instrument is set to off
•
The temperature display unit is set to °C
•
No automatic cell conditioning
•
No automatic cell levelling
•
No print of the measurement curve
•
No balance is connected
•
The default stirring is set to 10%
•
The user is set to "Default user"
25
Summary of Contents for TitraLab KF1000 Series
Page 1: ...DOC022 52 93082 TitraLab KF1000 series workstations User Manual 03 2019 Edition 9...
Page 2: ......
Page 6: ...Specifications 4...
Page 12: ...General information 10...
Page 20: ...Installation 18...
Page 30: ...Startup 28...
Page 40: ...Standard operations 38...
Page 54: ...Troubleshooting 52...
Page 60: ...Replacement parts and accessories 58...
Page 64: ...Change application parameters 62...
Page 66: ...Frequently asked questions 64...
Page 67: ......