28
SECTION 2,
continued
2.8.2
Wiring the Instrument
Connect the unpowered wires to the instrument power terminal as follows:
1. Strip the outside insulation of each wire back ¼-inch.
2. Connect the three wires to the terminal using the information in the
following table.
3. Ensure the voltage setting is correct and supply power to the instrument.
Figure 10
Power Connections
2.8.3
Voltage Selection for Alternate Voltage Operation
Note: The voltage selector switch
must be set properly for the line
voltage to be used. Improper setting
can result in serious damage to the
instrument when power is applied.
See Figure 11.
The instrument is switched for 115-volt operation when it leaves the factory.
All that is necessary to convert the instrument for 230-volt operation is to slide the
AC line conversion switch (shown in Figure 11) to the 230V position.
Wire Color Code for:
Protective Earth Ground
Hot or
ý
1
Neutral or
ý
2
North America
Green
Black
White
IEC
Green with Yellow Tracer
Brown
Blue
A.
B.
Summary of Contents for SP-510
Page 1: ...Cat No 54100 18 SP 510 HARDNESS MONITOR INSTRUMENT MANUAL...
Page 3: ...2 TRADEMARKS OF HACH COMPANY...
Page 13: ...12...
Page 17: ...16...
Page 20: ...SECTION 2 continued 19 Figure 3 SP 510 Instrument Dimensions 2 of 3...
Page 21: ...20 SECTION 2 continued Figure 4 SP 510 Instrument Dimensions 3 of 3...
Page 26: ...SECTION 2 continued 25 Figure 7 Sample Conditioning Kit...
Page 33: ...32 SECTION 2 continued Figure 15 Installing Pump Valve Module Pinch Plate...
Page 35: ...34...
Page 41: ...40...
Page 43: ...42...
Page 49: ...48...
Page 51: ...50...
Page 53: ...52...
Page 55: ...54 HOW TO ORDER...
Page 56: ...55 REPAIR SERVICE...
Page 60: ...59...