SPECIFICATIONS
GENERAL SPECIFICATIONS
Hard/Soft Display: Two LEDS with one indicating hard water and the second
one indicating soft water
Time Delay: A soft-to-hard water alarm transition requires two successive
readings above the trip point. A hard-to-soft water alarm clearing transition
requires one reading below the trip point.
Trip Points:
1.0, 2.0, 5.0, 10.0, 20.0, 50.0, and 100.0 mg/L hardness measured as CaCO
3
Reagents Required: Buffer and indicator for the desired trip point
Reagent Bottle Volume: 500 mL
Reagent Shelf Life: 1 year
Reagent Consumption Rate: 500 mL of each reagent every two months
Sampling Rate:
New sample every 1.9 minutes ± 5% at 60 Hz; new sample every 2.3 minutes
± 5% at 50 Hz
Controls: On/Off Switch, ALARM OFF, ZERO CAL, HARD CAL
Service Requirements:
Restandardization every 2 months. Replace pump tubes every 6 months.
Trip Point
Buffer Cat. No.
Indicator Cat. No.
1 mg/L
27685-49
27690-49
2 mg/L
27685-49
27691-49
5 mg/L
27685-49
27692-49
10 mg/L
27686-49
27692-49
20 mg/L
27687-49
27692-49
50 mg/L
27688-49
27692-49
100 mg/L
27689-49
27692-49
Summary of Contents for SP-510
Page 1: ...Cat No 54100 18 SP 510 HARDNESS MONITOR INSTRUMENT MANUAL...
Page 3: ...2 TRADEMARKS OF HACH COMPANY...
Page 13: ...12...
Page 17: ...16...
Page 20: ...SECTION 2 continued 19 Figure 3 SP 510 Instrument Dimensions 2 of 3...
Page 21: ...20 SECTION 2 continued Figure 4 SP 510 Instrument Dimensions 3 of 3...
Page 26: ...SECTION 2 continued 25 Figure 7 Sample Conditioning Kit...
Page 33: ...32 SECTION 2 continued Figure 15 Installing Pump Valve Module Pinch Plate...
Page 35: ...34...
Page 41: ...40...
Page 43: ...42...
Page 49: ...48...
Page 51: ...50...
Page 53: ...52...
Page 55: ...54 HOW TO ORDER...
Page 56: ...55 REPAIR SERVICE...
Page 60: ...59...