46
SECTION 5,
continued
5.2
Unscheduled Maintenance
5.2.1
Changing Instrument Alarm Point
Changing the instrument to alarm at a different hardness level is accomplished by
installing the appropriate buffer and indicator solutions, see Section 3.2 on
page 35. Following installation of the appropriate reagents, allow the instrument
to run for two hours to purge the old reagents. Recalibrate the instrument.
5.2.2
Fuse Replacement
The T, 1.25A, 250V fuse used in this instrument is used for both 115V and
230V operation.
DANGER
Remove power from the instrument
when removing or installing a
fuse.
PELIGRO
Apagar la electricidad del instrumento al quitar o instalar fusibles.
PERIGO
Desconecte a energia do instrumento ao remover ou instalar um fusível.
DANGER
Couper l'alimentation électrique de l'appareil pour retirer ou installer un fusible.
GEFAHR
Beim Entfernen oder Einsetzen der Sicherung muss die Stromzufuhr zum Gerät
unterbrochen werden.
1. Make sure there is no power supplied to the instrument.
2. Remove the customer access cover.
3. Locate the fuse holders (near the terminal strip in the customer wiring
compartment). See Figure 11 on page 29.
DANGER
For continued protection against
fire hazard, replace the fuse
only with a fuse of the same type
and rating.
PELIGRO
Para una continua protección contra incendios, reemplace los fusibles únicamente por
los del tipo y capacidad recomendados.
PERIGO
Para proteçao contínua contra fogo, troque os fusíveis somente por fusíveis do tipo
especificado para a força da corrente.
DANGER
Pour assurer la protection contre les risques d'incendies, remplacez les fusibles
uniquement par des fusibles du même type et pour la même intensité.
GEFAHR
Zur Wahrung des kontinuierlichen Brandschutzes dürfen die Sicherungen nur mit
Sicherungen des gleichen Typs und mit gleichen Stromkennwerten verwendet werden.
4. Remove the two fuses (F1 and F2) and replace them with two new fuses with
the same specifications (T, 1.25A, 250V). See REPLACEMENT ITEMS on
page 53.
5. Reinstall the customer access cover and resupply power.
Summary of Contents for SP-510
Page 1: ...Cat No 54100 18 SP 510 HARDNESS MONITOR INSTRUMENT MANUAL...
Page 3: ...2 TRADEMARKS OF HACH COMPANY...
Page 13: ...12...
Page 17: ...16...
Page 20: ...SECTION 2 continued 19 Figure 3 SP 510 Instrument Dimensions 2 of 3...
Page 21: ...20 SECTION 2 continued Figure 4 SP 510 Instrument Dimensions 3 of 3...
Page 26: ...SECTION 2 continued 25 Figure 7 Sample Conditioning Kit...
Page 33: ...32 SECTION 2 continued Figure 15 Installing Pump Valve Module Pinch Plate...
Page 35: ...34...
Page 41: ...40...
Page 43: ...42...
Page 49: ...48...
Page 51: ...50...
Page 53: ...52...
Page 55: ...54 HOW TO ORDER...
Page 56: ...55 REPAIR SERVICE...
Page 60: ...59...