9.
Review the calibration result:
• PASS—the sensor is calibrated and the calibration factor is displayed.
• FAIL—the calibration is outside of accepted limits. Repeat the calibration with fresh reference
on page 16 for more information.
10.
If the calibration passed, push
enter
to continue.
11.
If the option for operator ID is set to YES in the CAL OPTIONS menu, enter an operator ID.
12.
On the NEW SENSOR screen, select whether the sensor is new:
Option Description
YES
The sensor was not calibrated previously with this controller. The days of operation and previous
calibration curves for the sensor are reset.
NO
The sensor was calibrated previously with this controller.
13.
Return the sensor to the process and push
enter
.
The output signal returns to the active state and the measured sample value is shown on the
measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active
state.
Calibration with the process sample
The sensor can remain in the process sample.
1.
Push the
menu
key and select SENSOR SETUP>CALIBRATE.
2.
If the pass code is enabled in the security menu for the controller, enter the pass code.
3.
Select SAMPLE CAL and push
enter
.
4.
Select the option for the output signal during calibration:
Option
Description
ACTIVE
The instrument sends the current measured output value during the calibration procedure.
HOLD
The sensor output value is held at the current measured value during the calibration procedure.
TRANSFER
A preset output value is sent during calibration. Refer to the controller user manual to change
the preset value.
5.
With the sensor in the process sample, push
enter
. The measured value is shown. Wait for the
value to stabilize and push
enter
.
6.
With a certified secondary verification instrument measure the concentration value of the sample.
To avoid impurities in the sample take the measurement before the sample enters the flow
chamber. Use the arrow keys to enter this value if different from the value displayed and push
enter
.
7.
Review the calibration result:
• PASS—the sensor is calibrated and the calibration factor is displayed.
• FAIL—the calibration is outside of accepted limits. Clean the sensor and retry. Refer to
on page 16 for more information.
8.
If the calibration passed, push
enter
to continue.
9.
If the option for operator ID is set to YES in the CAL OPTIONS menu, enter an operator ID. Refer
to
on page 13.
10.
On the NEW SENSOR screen, select whether the sensor is new:
Option Description
YES
The sensor was not calibrated previously with this controller. The days of operation and previous
calibration curves for the sensor are reset.
NO
The sensor was calibrated previously with this controller.
12
English
Summary of Contents for Polymetron 9582sc
Page 26: ...26 Deutsch...
Page 45: ...Italiano 45...
Page 64: ...64 Fran ais...
Page 83: ...Espa ol 83...
Page 102: ...102 Portugu s...
Page 116: ...ESD 116...
Page 117: ...EMC A A A A EMC 117...
Page 118: ...MSDS SDS 1 1 2 5 mL 3 4 5 1 2 1 2 3 4 118...
Page 119: ...2 1 5 2 6 1 4 NPT 3 7 M4 4 8 1 4 NPT 119...
Page 120: ...1 1 1 TEMP 2 Temp 3 4 5 6 7 8 9 10 11 12 120...
Page 121: ...115 1 2 3 4 5 121...
Page 126: ...3 4 1 2 3 4 5 6 118 126...
Page 127: ...4 3 12 AgBr 90 2 3 N 400 600 30 127...
Page 128: ...6 C 43 F 35 C 95 F 131 AgBr N 400 600 128...
Page 129: ...EMC H2S 4 H2S H2S O2 4 AgBr N 400 600 0 ppb 10 000 ppb 2 ppm 129...
Page 156: ...156 Nederlands...
Page 193: ...Svenska 193...
Page 229: ...Magyar 229...
Page 243: ...243...
Page 244: ...EMC EMC A A 244...
Page 245: ...MSDS SDS 1 1 2 5 3 4 5 1 245...
Page 246: ...2 1 2 3 4 2 1 5 2 6 1 4 NPT 3 7 4 4 8 1 4 NPT 246...
Page 247: ...1 1 1 2 3 4 5 6 7 8 9 10 11 12 247...
Page 248: ...248...
Page 254: ...MSDS SDS 1 2 3 3 3 254...
Page 255: ...4 1 2 3 4 5 6 245 4 3 12 AgBr 90 2 3 N 400 600 30 255...
Page 256: ...6 C 43 F 35 C 95 F 260 256...
Page 257: ...AgBr N 400 600 H2S 4 257...
Page 258: ...H2S H2S O2 4 AgBr 400 600 0 10 000 2 258...
Page 267: ...T rk e 267...
Page 279: ......