Troubleshooting
General troubleshooting
Problem
Probable cause
Resolution
Excessive time for
stabilization or no
stabilization during
calibration in air.
The sample temperature is very different
from the ambient temperature, e.g. 6 °C
(43 °F) in water and 35 °C (95 °F) in air will
cause a measurement drift.
Do not wait till the sensor temperature
equals the external temperature (use
temperature compensation).
There is an electrolyte leak through the
membrane. The current is too high because
of an excessive penetration of oxygen.
Change the membrane.
There is pollution of the electrolyte due to a
loose filling screw.
Change the electrolyte. Check there is a
gasket in place and tighten the screw using
a screwdriver but without using excessive
force.
The electrode is not correctly mounted in
the sensor body causing an excessive gap
between the membrane and the cathode.
Tighten the electrode holding nut.
The membrane is incorrectly installed
causing a risk of electrolyte pollution.
Change the electrolyte and reinstall the
membrane onto the sensor body as far as it
will go finger tight.
Lack of humidity due to high temperature.
on page 19).
There is water or humidity in the sensor
cable connector.
Dry the sensor cable connector inside and
out, and install finger tight.
The gold surface is scratched or damaged.
Change the electrode.
Sludge or particles on the cathode.
Clean the cathode with a soft and
absorbent tissue. Rinse the membrane.
Cable or connections damaged when the
sensor has been removed.
Check the sensor connection to the
controller module. If correct check the
connections to the sensor cable connector.
The sensor is incorrectly positioned. The
electrolyte has leaked and air bubbles have
entered the cathode.
Put the sensor in the correct position, head
down.
No significant
current increase
when the sensor is
in the air for the
calibration.
There is pollution of the electrolyte due to a
loose filling screw.
Change the electrolyte. Check there is a
gasket in place and tighten the screw using
a screwdriver but without using excessive
force. Check the sensor is not damaged.
There is pollution of the electrolyte due to a
leaking membrane.
Change the electrolyte and the membrane.
The membrane is torn.
Change the membrane.
The electrode is not correctly mounted in
the sensor body causing an excessive gap
between the membrane and the cathode.
Tighten the electrode holding nut.
Cable or connections damaged when the
sensor has been removed.
Check the connection to the controller
module. If correct check the connections to
the sensor cable connector.
The membrane is worn.
Change the membrane.
There is a dark green deposit of silver
bromide (AgBr) on the silver tube.
Polish the tube with a soft abrasive (N°
400 to 600) and replace the membrane.
16
English
Summary of Contents for Polymetron 9582sc
Page 26: ...26 Deutsch...
Page 45: ...Italiano 45...
Page 64: ...64 Fran ais...
Page 83: ...Espa ol 83...
Page 102: ...102 Portugu s...
Page 116: ...ESD 116...
Page 117: ...EMC A A A A EMC 117...
Page 118: ...MSDS SDS 1 1 2 5 mL 3 4 5 1 2 1 2 3 4 118...
Page 119: ...2 1 5 2 6 1 4 NPT 3 7 M4 4 8 1 4 NPT 119...
Page 120: ...1 1 1 TEMP 2 Temp 3 4 5 6 7 8 9 10 11 12 120...
Page 121: ...115 1 2 3 4 5 121...
Page 126: ...3 4 1 2 3 4 5 6 118 126...
Page 127: ...4 3 12 AgBr 90 2 3 N 400 600 30 127...
Page 128: ...6 C 43 F 35 C 95 F 131 AgBr N 400 600 128...
Page 129: ...EMC H2S 4 H2S H2S O2 4 AgBr N 400 600 0 ppb 10 000 ppb 2 ppm 129...
Page 156: ...156 Nederlands...
Page 193: ...Svenska 193...
Page 229: ...Magyar 229...
Page 243: ...243...
Page 244: ...EMC EMC A A 244...
Page 245: ...MSDS SDS 1 1 2 5 3 4 5 1 245...
Page 246: ...2 1 2 3 4 2 1 5 2 6 1 4 NPT 3 7 4 4 8 1 4 NPT 246...
Page 247: ...1 1 1 2 3 4 5 6 7 8 9 10 11 12 247...
Page 248: ...248...
Page 254: ...MSDS SDS 1 2 3 3 3 254...
Page 255: ...4 1 2 3 4 5 6 245 4 3 12 AgBr 90 2 3 N 400 600 30 255...
Page 256: ...6 C 43 F 35 C 95 F 260 256...
Page 257: ...AgBr N 400 600 H2S 4 257...
Page 258: ...H2S H2S O2 4 AgBr 400 600 0 10 000 2 258...
Page 267: ...T rk e 267...
Page 279: ......