background image

Troubleshooting

General troubleshooting

Problem

Probable cause

Resolution

Excessive time for
stabilization or no
stabilization during
calibration in air.

The sample temperature is very different
from the ambient temperature, e.g. 6 °C
(43 °F) in water and 35 °C (95 °F) in air will
cause a measurement drift.

Do not wait till the sensor temperature
equals the external temperature (use
temperature compensation).

There is an electrolyte leak through the
membrane. The current is too high because
of an excessive penetration of oxygen.

Change the membrane.

There is pollution of the electrolyte due to a
loose filling screw.

Change the electrolyte. Check there is a
gasket in place and tighten the screw using
a screwdriver but without using excessive
force.

The electrode is not correctly mounted in
the sensor body causing an excessive gap
between the membrane and the cathode.

Tighten the electrode holding nut.

The membrane is incorrectly installed
causing a risk of electrolyte pollution.

Change the electrolyte and reinstall the
membrane onto the sensor body as far as it
will go finger tight.

Lack of humidity due to high temperature.

Use a calibration cap (see 

Replacement

parts and accessories

 on page 19).

There is water or humidity in the sensor
cable connector.

Dry the sensor cable connector inside and
out, and install finger tight.

The gold surface is scratched or damaged.

Change the electrode.

Sludge or particles on the cathode.

Clean the cathode with a soft and
absorbent tissue. Rinse the membrane.

Cable or connections damaged when the
sensor has been removed.

Check the sensor connection to the
controller module. If correct check the
connections to the sensor cable connector.

The sensor is incorrectly positioned. The
electrolyte has leaked and air bubbles have
entered the cathode.

Put the sensor in the correct position, head
down.

No significant
current increase
when the sensor is
in the air for the
calibration.

There is pollution of the electrolyte due to a
loose filling screw.

Change the electrolyte. Check there is a
gasket in place and tighten the screw using
a screwdriver but without using excessive
force. Check the sensor is not damaged.

There is pollution of the electrolyte due to a
leaking membrane.

Change the electrolyte and the membrane.

The membrane is torn.

Change the membrane.

The electrode is not correctly mounted in
the sensor body causing an excessive gap
between the membrane and the cathode.

Tighten the electrode holding nut.

Cable or connections damaged when the
sensor has been removed.

Check the connection to the controller
module. If correct check the connections to
the sensor cable connector.

The membrane is worn.

Change the membrane.

There is a dark green deposit of silver
bromide (AgBr) on the silver tube.

Polish the tube with a soft abrasive (N°
400 to 600) and replace the membrane.

16

   

English

Summary of Contents for Polymetron 9582sc

Page 1: ...Benutzerhandbuch Manuale di base per l utente Manuel d utilisation de base Manual b sico del usuario Manual b sico do utilizador Z kladn u ivatelsk p ru ka Basisgebruikershandleiding Grundl ggende bru...

Page 2: ...English 3 Deutsch 20 Italiano 39 Fran ais 58 Espa ol 77 Portugu s 96 115 e tina 132 Nederlands 150 Dansk 169 Svenska 187 Suomi 206 Magyar 223 242 T rk e 261 2...

Page 3: ...cification Details Sensor body Noryl Sensor body with optional immersion Stainless steel 316L Dimensions 144 x 144 x 181 mm 5 7 x 5 7 x 7 1 in Weight 1 7 kg 3 75 lb Electrode Cathode gold Anode silver...

Page 4: ...a potentially or imminently hazardous situation which if not avoided will result in death or serious injury W A R N I N G Indicates a potentially or imminently hazardous situation which if not avoide...

Page 5: ...heat transfer liquid Product components Make sure that all components have been received If any items are missing or damaged contact the manufacturer or a sales representative immediately Components...

Page 6: ...ll and tighten the holding nut by hand 5 Install the gasket and filling screw to prevent electrolyte leakage or external contamination from the sample Carefully tighten the screw with a screwdriver wi...

Page 7: ...g and wiring instructions Connect the sensor to the controller W A R N I N G Potential Electrocution Hazard Always disconnect power to the instrument when making electrical connections W A R N I N G E...

Page 8: ...e 1 Be sure to connect all sensor ground shield wires to the controller enclosure grounding screws Note If the sensor cable is not long enough to reach the controller an interconnect cable and junctio...

Page 9: ...3 Examine the plumbing for leaks and stop any leaks if found 4 Apply power to the controller 5 Make the applicable menu selections when the controller starts User navigation Refer to the controller do...

Page 10: ...erval for data storage in the data log 5 30 seconds 1 2 5 10 15 default 30 60 minutes RESET DEFAULTS Sets the configuration menu to the default settings All sensor information is lost Calibrate the se...

Page 11: ...or the sensor are reset NO The sensor was calibrated previously with this controller 12 Return the sensor to the process and push enter The output signal returns to the active state and the measured s...

Page 12: ...ibration Option Description ACTIVE The instrument sends the current measured output value during the calibration procedure HOLD The sensor output value is held at the current measured value during the...

Page 13: ...elect SENSOR SETUP CALIBRATE 2 If the pass code is enabled in the security menu for the controller enter the pass code 3 Select CAL OPTIONS and push enter 4 Use the arrow keys to select an option and...

Page 14: ...the sensor clamping nut from the flow chamber and remove the sensor Figure 3 Figure 3 Remove the sensor Replace the membrane Refer to the steps that follow and Figure 4 to replace the sensor membrane...

Page 15: ...and membrane visually check the silver anode If more than 2 3 of the surface is coated with silver bromide then an electrode rejuvenation is required To do this very softly polish the areas coated wit...

Page 16: ...all finger tight The gold surface is scratched or damaged Change the electrode Sludge or particles on the cathode Clean the cathode with a soft and absorbent tissue Rinse the membrane Cable or connect...

Page 17: ...membrane filling screw and change the electrolyte and the membrane Electrolyte leaking Check the screw in parts membrane filling screw and change the electrolyte and the membrane Interferences mainly...

Page 18: ...ial number for the sensor module SENSOR INFORMATION Shows the name and serial number that was entered by the user CAL DAYS Shows the number of days since the last calibration CAL HISTORY Shows a list...

Page 19: ...rane days in the DIAG TEST menu NOT CALIBRATED The sensor has not been calibrated Calibrate the sensor CAL IN PROGRESS A calibration was started but not completed Return to calibration Replacement par...

Page 20: ...g nderungen unterliegen Sensor Spezifikation Details Sensorgeh use Noryl Sensorgeh use ohne optionale Immersion Edelstahl 316L Abmessungen 144 x 144 x 181 mm 5 7 x 5 7 x 7 1 Zoll Gewicht 1 7 kg 3 75 l...

Page 21: ...ystem nur wie in diesem Handbuch beschrieben Bedeutung von Gefahrenhinweisen G E F A H R Kennzeichnet eine m gliche oder drohende Gefahrensituation die wenn sie nicht vermieden wird zum Tod oder zu sc...

Page 22: ...ng von gel stem Sauerstoff in Kesselspeisewasser Vorw rmern Kondensatoren und ganz allgemein in allen thermischen Anlagen die Wasser f r die W rme bertragung einsetzen verwendet Produktkomponenten Ste...

Page 23: ...nden um 5 ml Elektrolyt zum Sensorgeh use hinzuzuf gen Sicherstellen dass der Elektrolyt frei von Blasen und Verunreinigungen ist 3 Die Elektrode vorsichtig in das Sensorgeh use dr cken bis ein Widers...

Page 24: ...an 4 Schlie en Sie den Probenausgang an den Probenausgangsanschluss an Abbildung 2 Sensor montieren 1 Kabelanschluss Sensor 5 Flusskammer 2 Spannmutter Sensor 6 1 4 NPT Probeneinlassanschluss nicht i...

Page 25: ...I S M glicher Ger teschaden Empfindliche interne elektronische Bauteile k nnen durch statische Elektrizit t besch digt werden wobei dann das Ger t mit verminderter Leistung funktioniert oder schlie li...

Page 26: ...26 Deutsch...

Page 27: ...Zahlen Zeichen oder Satzzeichen die Pfeiltasten nach oben oder nach unten dr cken und halten Mit der rechten Pfeiltaste zum n chsten Feld gehen Option Beschreibung EDIT NAME Definiert den dem Sensor...

Page 28: ...zeigt Null Kalibrierung Aufgrund der Stabilit t der Elektroden ist f r die meisten Anwendungen keine Kalibrierung erforderlich Sie kann aber verwendet werden um den eindeutigen Nullpunkt des Sensors z...

Page 29: ...ie AIR CAL Luftkalibrierung und dr cken Sie Enter 5 W hlen Sie die Option f r das Ausgangssignal w hrend der Kalibrierung Option Beschreibung ACTIVE Das Ger t bermittelt w hrend der Kalibrierung den a...

Page 30: ...librierung einen vorher eingestellten Ausgangswert F r die nderung des voreingestellten Wertes beziehen Sie sich bitte auf das Benutzerhandbuch des Controllers 5 Mit dem Sensor in der Prozessprobe Ent...

Page 31: ...e Kalibrierungserinnerung einstellen oder eine Bediener ID mit Kalibrierungsdaten aus diesem Men einf gen 1 Die Men Taste dr cken und SENSOR SETUP CALIBRATE Sensoreinstellung Kalibrieren ausw hlen 2 W...

Page 32: ...n mit Verschlei erscheinungen Bauen Sie den Sensor aus 1 Probenzulauf ausschalten 2 Sensorkabelanschluss von dem Sensor abschrauben Nicht direkt am Kabel oder am Sensor drehen 3 Die Spannmutter des Se...

Page 33: ...kt sind Beim Austausch des Elektrolyten und der Membran nehmen Sie immer eine Sichtkontrolle der Silberanode vor Wenn mehr als 2 3 der Fl che mit Silberbromid bedeckt sind m ssen die Elektroden erneue...

Page 34: ...Membran und Kathode Ziehen Sie die Haltemutter der Elektrode an Die Membran wurde nicht korrekt installiert Deshalb k nnen Verunreinigungen in den Elektrolyt eindringen Tauschen Sie den Elektrolyt aus...

Page 35: ...Silberrohr sind dunkelgr ne Ablagerungen aus Silberbromid AgBr Reinigen Sie das Rohr mit einem Schleifmittel mit feiner K rnung P400 bis P600 und tauschen Sie die Membran aus Auff llige Schwankungen i...

Page 36: ...n Flussrate zu niedrig mindestens 4 ml h Probenfluss erh hen Die Probentemperatur oder der Druck liegen au erhalb der Spezifikation ndern Sie die Position des Sensors oder die Probe damit die Werte in...

Page 37: ...ehen verloren SENSORSIGNALE Zeigt den aktuellen Messwert in mV und die Temperatur an MEMBRANEN TAGE Zeigt die Anzahl der Tage an die der Sensor in Betrieb gewesen ist MEMBR ZUR CKS Setzt die Anzahl de...

Page 38: ...r wurde nicht kalibriert Kalibrieren Sie den Sensor CAL IN PROGRESS Eine Kalibrierung wurde gestartet aber noch nicht abgeschlossen Kehren Sie zur Kalibrierung zur ck Ersatzteile und Zubeh r F r Infor...

Page 39: ...gli Corpo del sensore Noryl Corpo del sensore con immersione opzionale Acciaio inossidabile 316L Dimensioni 144 x 144 x 181 mm 5 7 x 5 7 x 7 1 pollici Peso 1 7 kg 3 75 libbre Elettrodo Catodo oro Anod...

Page 40: ...e durante l utilizzo dello stesso Non utilizzare o installare questa apparecchiatura in modo diverso da quanto specificato nel presente manuale Indicazioni e significato dei segnali di pericolo P E R...

Page 41: ...un canale utilizzato per la misurazione dell ossigeno disciolto nella acque di alimentazione delle caldaie in economizzatori condensatori e in generale in tutte le apparecchiature che utilizzano l ac...

Page 42: ...fondo 2 Utilizzare la siringa per aggiungere 5 ml di elettrolita al corpo del sensore Assicurarsi che non vi siano impurit o bolle nell elettrolita 3 Premere delicatamente l elettrodo nel corpo del se...

Page 43: ...re l uscita del campione al connettore di uscita Figura 2 Installazione del sensore 1 Connettore del cavo del sensore 5 Cella di flusso 2 Dado di serraggio del sensore 6 Connettore ingresso campione d...

Page 44: ...analogiche A V V I S O Danno potenziale allo strumento Componenti elettronici interni delicati possono essere danneggiati dall elettricit statica compromettendo le prestazioni o provocando guasti Far...

Page 45: ...Italiano 45...

Page 46: ...tura tenere premuto il tasto freccia su o gi Premere il tasto freccia destra per spostarsi allo spazio successivo Opzione Descrizione MODIFICA NOME Consente di modificare il nome del sensore visualizz...

Page 47: ...o assoluto del sensore 1 Rimuovere il sensore dal processo e sciacquarlo in acqua distillata 2 Premere il tasto menu e selezionare SENSOR SETUP Configurazione sensore CALIBRATE Calibra 3 Se stata impo...

Page 48: ...predefinito viene inviato durante la calibrazione Per modificare il valore predefinito fare riferimento al manuale dell utente del controller 6 Utilizzando un barometro di precisione certificato misur...

Page 49: ...del campione utilizzando un secondo strumento di verifica certificato Per evitare impurit nel campione eseguire la misurazione prima che il campione entri nella cella di flusso Utilizzare i tasti frec...

Page 50: ...invio 4 Utilizzare i tasti freccia per selezionare un opzione e premere invio Opzione Descrizione PROMEM CALIBRAZIONE Imposta un promemoria per la calibrazione successiva in giorni mesi o anni Selezio...

Page 51: ...cella di flusso e rimuovere il sensore Figura 3 Figura 3 Rimuovere il sensore Sostituzione della membrana Fare riferimento ai passaggi seguenti e alla Figura 4 per sostituire la membrana del sensore 1...

Page 52: ...della membrana controllare l anodo in argento Se oltre i 2 3 della superficie sono ricoperti di bromuro d argento procedere con il rinnovamento del sensore Lucidare delicamente le aree interessate us...

Page 53: ...tringere il dado di trattenimento dell elettrodo La membrana non installata correttamente ed espone l elettrolita al rischio di inquinamento Sostituire l elettrolita e reinstallare la membrana sul cor...

Page 54: ...i formato un deposito verde scuro di bromuro d argento AgBr sul tubo in argento Lucidare il tubo con un leggero abrasivo N da 400 a 600 e sostituire la membrana Grave instabilit nel modo di misurazion...

Page 55: ...one del sensore oppure sostituire il campione in modo che rientri nelle specifiche Depositi di sostanze o particelle sul catodo Pulire il catodo con un panno morbido e assorbente Sostituire la membran...

Page 56: ...rmata principale sotto il valore di misurazione Un messaggio di avvertenza non influenza il funzionamento di menu rel e uscite Per visualizzare i messaggi di avviso premere il tasto menu e selezionare...

Page 57: ...ller consultare la sezione Parti di ricambio e accessori della documentazione del controller Nota Prodotti e numeri articolo possono variare in funzione del paese di commercializzazione Contattare il...

Page 58: ...ification D tails Corps du capteur Noryl Corps du capteur avec immersion en option Acier inoxydable 316L Dimensions 144 x 144 x 181 mm 5 7 x 5 7 x 7 1 po Poids 1 7 kg 3 75 lb lectrode Cathode or Anode...

Page 59: ...d crite dans ce manuel Interpr tation des indications de risques D A N G E R Indique une situation de danger potentiel ou imminent qui si elle n est pas vit e entra ne des blessures graves voire mort...

Page 60: ...eaux d alimentation de chaudi res les conomiseurs les condensateurs et en g n ral dans tout quipement thermique utilisant l eau comme fluide de transfert de chaleur Composants du produit Assurez vous...

Page 61: ...d lectrolyte dans le corps du capteur V rifiez l absence d impuret s ou de bulles dans l lectrolyte 3 Poussez d licatement l lectrode dans le corps du capteur jusqu la but e Ne forcez pas sur l lectro...

Page 62: ...z la sortie d chantillon au connecteur de sortie d chantillon Figure 2 Montage du capteur 1 Connecteur pour c ble de capteur 5 Chambre de circulation 2 crou de serrage du capteur 6 Connecteur d entr e...

Page 63: ...V I S D g t potentiel sur l appareil Les composants lectroniques internes de l appareil peuvent tre endommag s par l lectricit statique qui risque d alt rer ses performances et son fonctionnement Repo...

Page 64: ...64 Fran ais...

Page 65: ...ONFIG CAPTEUR CONFIGURER 2 Utilisez les touches fl ch es pour s lectionner une option et appuyez sur entr e Pour saisir les num ros les caract res ou la ponctuation appuyez et maintenez enfonc les tou...

Page 66: ...affich e et appuyez sur entr e 7 Appuyez sur entr e pour confirmer l talonnage Le d calage de temp rature est affich talonnage du z ro En raison de la stabilit de l lectrode pour la plupart des appli...

Page 67: ...valeur de sortie du capteur est maintenue la derni re valeur mesur e pendant la proc dure d talonnage TRANSFER Transfert Une valeur de sortie pr d finie est envoy e pendant l talonnage Consultez le ma...

Page 68: ...l chantillon n entre dans la chambre de circulation Utilisez les touches fl ch es pour saisir cette valeur si elle est diff rente de la valeur affich e et appuyez sur entr e 7 Consultez le r sultat d...

Page 69: ...UR TAL Inclut un identifiant op rateur avec les donn es d talonnage OUI ou NON par d faut L identifiant est saisi pendant l talonnage Fonction veille La plage de mesure de l instrument 9582 est de 2 p...

Page 70: ...etirez le capteur Remplacement de la membrane Reportez vous aux tapes suivantes et la Figure 4 pour remplacer la membrane du capteur 1 D vissez la vis de remplissage et retirez la avec le joint 2 D vi...

Page 71: ...contr le visuel de l anode en argent Si plus des 2 3 de la surface sont recouverts de bromure d argent la r g n ration de l lectrode est n cessaire Pour ce faire polissez tr s l g rement les zones re...

Page 72: ...brane et la cathode Serrez l crou de retenue de l lectrode La membrane est mal install e en cause un risque de pollution de l lectrolyte Remplacez l lectrolyte et r installez la membrane fond sur le c...

Page 73: ...onc de bromure d argent AgBr Polissez le tube avec un abrasif doux n 400 600 et remplacez la membrane Instabilit importante en mode mesure Il y a de l eau ou de l humidit dans le connecteur du c ble d...

Page 74: ...te qu il rentre dans les sp cifications Boue ou particules sur la cathode Nettoyez la cathode avec un chiffon doux et absorbant Remplacez la membrane Le courant du capteur est nul durant la mesure Abs...

Page 75: ...nes d avertissement apparaissent droite de l cran principal au dessous de la valeur de mesure Un avertissement n affecte pas le fonctionnement des menus relais et sorties Pour voir les avertissements...

Page 76: ...ation du contr leur pour conna tre les pi ces de rechange et les accessoires Remarque Les num ros de r f rence de produit et d article peuvent d pendre des r gions de commercialisation Prenez contact...

Page 77: ...aviso Sensor Especificaci n Detalles Cuerpo del sensor Noryl Cuerpo del sensor con inmersi n opcional Acero inoxidable 316L Dimensiones 144 x 144 x 181 mm 5 7 x 5 7 x 7 1 pulg Peso 1 7 kg 3 75 libras...

Page 78: ...sobre riesgos P E L I G R O Indica una situaci n potencial o de riesgo inminente que de no evitarse provocar la muerte o lesiones graves A D V E R T E N C I A Indica una situaci n potencial o inminent...

Page 79: ...ucto Este analizador de un solo canal sirve para la medici n del ox geno disuelto en aguas de abastecimiento de calderas economizadores condensadores y en general todos los equipos t rmicos que usan a...

Page 80: ...e llegue al tope 2 Utilice la jeringa para a adir 5 ml del electrolito al cuerpo del sensor Aseg rese de que no hay impurezas ni burbujas en el electrolito 3 Empuje suavemente el electrodo hacia el cu...

Page 81: ...alida de muestra en el conector de salida de muestra Figura 2 Montaje del sensor 1 Conector del cable del sensor 5 C mara de flujo 2 Tuerca de sujeci n del sensor 6 Conector de entrada de muestra de N...

Page 82: ...al al instrumento Los delicados componentes electr nicos internos pueden sufrir da os debido a la electricidad est tica lo que acarrea una disminuci n del rendimiento del instrumento y posibles fallos...

Page 83: ...Espa ol 83...

Page 84: ...ara seleccionar una opci n y pulse enter Intro Para introducir n meros caracteres o signos de puntuaci n pulse y mantenga pulsadas las teclas de flecha arriba y abajo Pulse la tecla de flecha derecha...

Page 85: ...ccione 1 PT TEMP CAL CALIBRACI N DE TEMPERATURA DE 1 PUNTO y pulse enter Intro 5 Aparece el valor de temperatura bruto Pulse enter Intro 6 Introduzca el valor correcto si es diferente del valor mostra...

Page 86: ...TUP CONFIGURACI N DE SENSOR CALIBRATE CALIBRAR 3 Si la contrase a de f brica est habilitada en el men de seguridad del controlador introduzca la contrase a 4 Seleccione AIR CAL CALIBRACI N AIRE y puls...

Page 87: ...lor de salida del sensor se mantiene en el valor de medici n actual TRANSFER TRANSFERIR Durante la calibraci n se env a un valor de salida predeterminado Consulte el manual del usuario del controlador...

Page 88: ...o de calibraci n o incluir un ID de operador con datos de calibraci n desde este men 1 Pulse la tecla menu Men y seleccione SENSOR SETUP CONFIGURACI N DE SENSOR CALIBRATE CALIBRAR 2 Si la contrase a d...

Page 89: ...utilice una membrana gastada Extracci n del sensor 1 Desconecte el suministro de la muestra 2 Desatornille el conector del cable del sensor del sensor No gure el cable o el sensor 3 Desatornille la tu...

Page 90: ...icie est contaminada Al cambiar el electrolito y la membrana examine visualmente el anodo de plata Si m s de 2 3 de la superficie est cubierta con bromuro de plata se necesita un rejuvenecimiento de l...

Page 91: ...vo entre la membrana y el c todo Apriete la tuerca de sujeci n del electrodo La membrana est instalada incorrectamente y causa el riesgo de contaminaci n del electrolito Cambie el electrolito y vuelva...

Page 92: ...bromuro de plata AgBr en el tubo de plata Limpie el tubo con un abrasivo suave n 400 a 600 y sustituya la membrana Inestabilidad importante en el modo de medici n Hay agua o humedad en el conector del...

Page 93: ...stra de modo que se halle dentro de las especificaciones Residuos o part culas sobre el c todo Limpie el c todo con un pa uelo suave y absorbente Cambie la membrana La corriente del sensor es nula dur...

Page 94: ...n Una advertencia no afecta el funcionamiento de los men s rel s y salidas Para ver las advertencias pulse la tecla menu Men y seleccione DIAGNOSTICS DIAGN STICO A continuaci n seleccione el dispositi...

Page 95: ...accesorios de la documentaci n del controlador para obtener informaci n sobre las piezas y los accesorios del controlador Nota Los n meros de producto y art culo pueden variar en algunas regiones de...

Page 96: ...Detalhes Corpo do sensor Noryl Corpo do sensor com imers o opcional A o inoxid vel 316L Dimens es 144 x 144 x 181 mm 5 7 x 5 7 x 7 1 pol Peso 1 7 kg El ctrodo C todo dourado nodo prateado Membrana PFA...

Page 97: ...r este equipamento n o comprometida N o o utilize ou instale sen o da forma especificada neste manual Uso da informa o de perigo P E R I G O Indica uma situa o de risco potencial ou eminente que se n...

Page 98: ...izado para a medi o do oxig nio dissolvido em guas de abastecimento de caldeiras economizadores condensadores e no geral em todo o equipamento t rmico que utilize gua como l quido de transfer ncia de...

Page 99: ...to ao corpo do sensor Certifique se de que n o existem impurezas ou bolhas de ar no electr lito 3 Introduza cuidadosamente o el ctrodo no interior do corpo do sensor at sentir resist ncia N o force o...

Page 100: ...ue a sa da da amostra ao conector de sa da da amostra Figura 2 Montagem do sensor 1 Conector do cabo do sensor 5 C mara de Fluxo 2 Porca de fixa o do sensor 6 Conector de entrada de amostra 1 4 NPT n...

Page 101: ...nto potencial Os componentes electr nicos internos sens veis podem ser danificados atrav s de electricidade est tica provocando um desempenho reduzido ou uma eventual falha Consulte os seguintes passo...

Page 102: ...102 Portugu s...

Page 103: ...e seta para seleccionar uma op o e prima enter Para introduzir n meros caracteres ou pontua o mantenha premidas as teclas de seta para cima ou para baixo Prima a tecla de seta para a direita para avan...

Page 104: ...1 ponto e prima enter 5 apresentado o valor bruto da temperatura Prima enter 6 Introduza o valor correcto se diferir do valor apresentado e prima enter 7 Prima enter para confirmar a calibra o apresen...

Page 105: ...guran a para o controlador introduza o c digo de acesso 4 Seleccione AIR CAL Calibra o com ar e prima enter 5 Seleccione a op o para o sinal de sa da durante a calibra o Op o Descri o ACTIVE Activo O...

Page 106: ...alibra o TRANSFER Transferir Um valor de sa da predefinido enviado durante a calibra o Consulte o manual do utilizador do controlador para alterar o valor predefinido 5 Com o sensor na amostra do proc...

Page 107: ...cluir uma ID de operador com dados de calibra o neste menu 1 Prima a tecla menu e seleccione SENSOR SETUP Configura o do sensor CALIBRATE Calibrar 2 Se o c digo de acesso estiver activado no menu de s...

Page 108: ...N o utilize uma membrana usada Remo o do sensor 1 Desligue o fornecimento de amostra 2 Desaperte o conector do cabo do sensor no sensor N o rode o pr prio cabo nem o sensor 3 Desaparafuse a porca da...

Page 109: ...iver contaminada Quando substituir o electr lito e a membrana realize uma verifica o visual do nodo de prata Caso mais de 2 3 da superf cie estejam cobertos por brometo de prata necess ria uma renova...

Page 110: ...ntre a membrana e o c todo Aperte a porca de reten o do el ctrodo A membrana est incorrectamente instalada o que provoca um risco de contamina o do electr lito Substitua o electr lito e volte a instal...

Page 111: ...e um dep sito verde escuro de brometo de prata AgBr no tubo de prata Efectue o polimento do tubo com um abrasivo suave N 400 a 600 e substitua a membrana Instabilidade significativa no modo de medi o...

Page 112: ...difique a amostra para que esteja dentro das especifica es Lodo ou part culas no c todo Limpe o c todo utilizando um len o de papel suave e absorvente Substitua a membrana A corrente do sensor nula du...

Page 113: ...s o apresentado direita do ecr principal sob o valor de medi o Um aviso n o afecta o funcionamento dos menus rel s e sa das Para visualizar erros prima a tecla menu e seleccione DIAGNOSTICS Diagn stic...

Page 114: ...de substitui o e acess rios da documenta o do controlador para obter as pe as e acess rios do controlador Nota Os n meros do produto e do artigo podem variar nalgumas regi es de venda Para mais infor...

Page 115: ...117 128 121 131 121 Noryl 316L 144 x 144 x 181 mm 5 7 x 5 7 x 7 1 in 1 7 kg 3 75 lb PFA Noryl 4 10 20 60 C 4 120 F 0 90 EN 61326 1 2006 EN 61010 1 2010 0 2 ppm 0 5 ppb 0 5 ppb 2 1 ppb 1 40 ppb 30 0 4...

Page 116: ...ESD 116...

Page 117: ...EMC A A A A EMC 117...

Page 118: ...MSDS SDS 1 1 2 5 mL 3 4 5 1 2 1 2 3 4 118...

Page 119: ...2 1 5 2 6 1 4 NPT 3 7 M4 4 8 1 4 NPT 119...

Page 120: ...1 1 1 TEMP 2 Temp 3 4 5 6 7 8 9 10 11 12 120...

Page 121: ...115 1 2 3 4 5 121...

Page 122: ...Enter EDIT NAME 16 16 MEAS UNITS PRESSURE UNITS TEMP UNITS C F FILTER 0 60 60 LOG SETUP 5 30 1 2 5 10 15 30 60 1 2 SENSOR SETUP CALIBRATE 3 4 1 PT TEMP CAL 1 Enter 5 Enter 6 Enter 7 Enter 1 2 SENSOR S...

Page 123: ...7 Enter 8 PASS FAIL 128 9 Enter 10 CAL OPTIONS ID YES ID 125 11 NEW SENSOR 12 Enter 1 2 SENSOR SETUP CALIBRATE 3 4 AIR CAL Enter 5 ACTIVE HOLD TRANSFER 6 Enter 7 Enter 8 Enter 9 PASS FAIL 128 10 Ente...

Page 124: ...12 NEW SENSOR 13 Enter 1 SENSOR SETUP CALIBRATE 2 3 SAMPLE CAL Enter 4 ACTIVE HOLD TRANSFER 5 Enter Enter 6 Enter 7 PASS FAIL 128 8 Enter 9 CAL OPTIONS ID YES ID 125 10 NEW SENSOR 11 Enter 1 124...

Page 125: ...BACK TO CAL SENSOR SETUP 2 Eter ID 1 SENSOR SETUP CALIBRATE 2 3 CAL OPTIONS Enter 4 Enter ID ID YES NO ID 9582 2 ppm 2 5 SENSOR SETUP EXIT STANDBY MSDS SDS 1 2 3 3 125...

Page 126: ...3 4 1 2 3 4 5 6 118 126...

Page 127: ...4 3 12 AgBr 90 2 3 N 400 600 30 127...

Page 128: ...6 C 43 F 35 C 95 F 131 AgBr N 400 600 128...

Page 129: ...EMC H2S 4 H2S H2S O2 4 AgBr N 400 600 0 ppb 10 000 ppb 2 ppm 129...

Page 130: ...MODULE INFORMATION SENSOR INFORMATION CAL DAYS CAL HISTORY RESET CAL HISTORY SENSOR SIGNALS mV MEMBRANE DAYS RESET MEMBRANE DIAGNOSTICS 3 3 DO 40 ppm DO DO 0 ppb TEMP TOO HIGH 50 C TEMP TOO LOW 0 C 20...

Page 131: ...4 09185 A 3500 09182 A 1200 09078 C 1020 09078 C 1030 ppb 09182 A 1000 ppb 09078 C 1010 25 09181 A 3600 09078 A 2000 460150 21951 131...

Page 132: ...parametry Podrobnosti T leso senzoru Noryl T lo sn ma e s voliteln m pono en m Nerezov ocel 316L Rozm ry 144 x 144 x 181 mm 5 7 x 5 7 x 7 1 Hmotnost 1 7 kg 3 75 libry Elektroda Katoda zlato anoda st b...

Page 133: ...einstalujte jinak ne jak je uvedeno v tomto n vodu Informace o mo n m nebezpe N E B E Z P E Ozna uje mo nou nebo bezprost edn rizikovou situaci je m e v p pad e j nezabr n te v st k usmrcen nebo v n m...

Page 134: ...u t n ho kysl ku v nap jec vod kotl ekonomiz r kondenz tor a obecn v ech tepeln ch za zen kter pou vaj vodu jako kapaliny pro p enos tepla Sou sti v robku Ujist te se e byly dod ny v echny sou sti V p...

Page 135: ...nepoc t te odpor Elektrodu nezatla ujte silou 4 Rukou na roubujte a ut hn te p dr nou matici 5 Na roubujte plnic roub s t sn n m aby nemohlo doj t k niku elektrolytu ani k jeho zne i t n vzorkem roub...

Page 136: ...naleznete v dokumentaci ke kontrol ru P ipojte sondu ke kontrol ru V A R O V N Nebezpe poran n el proudem P ed elektrick mi instalacemi p stroj odpojte od elektrick s t V A R O V N Nebezpe smrteln ho...

Page 137: ...Tabulka 1 Ujist te se e jsou v echny zemnic st nic vodi e senzoru p ipojeny k zemnic m roub m pouzdra kontrol ru Pozn mka Pokud nen kabel senzoru dostate n dlouh na to aby dos hl ke kontrol ru pro pr...

Page 138: ...kontrolujte zda na potrub nedoch z k nik m a odstra te p iny jak chkoliv zji t n ch nik 4 Zave te proud do kontrol ru 5 P i spu t n kontrol ru prove te p slu n volby v nab dce Navigace u ivatele Dal i...

Page 139: ...z a 60 sekund pr m rn hodnota sign lu po dobu 60 sekund Filtr zvy uje dobu po kterou m e sign l sn ma e reagovat na aktu ln zm ny v procesu LOG SETUP Nastaven logov n Nastav asov interval pro uchov n...

Page 140: ...opakujte s erstv m referen n m roztokem Dal informace naleznete v e en probl m na stran 145 9 Pokud byla kalibrace v po dku a chcete pokra ovat stiskn te enter 10 Pokud je mo nost pro ID obsluhy v nab...

Page 141: ...n k ivky sn ma e jsou resetov ny NE Sn ma byl d ve kalibrov n s t mto adi em 13 Vra te sn ma zp t do provozu a stiskn te enter V choz sign l se vr t do aktivn ho stavu a na obrazovce m en se zobraz na...

Page 142: ...brace bude zastavena Nov kalibrace mus za t zase od za tku BACK TO CAL Zp t ke kalibraci N vrat ke kalibraci LEAVE CAL Zastavit kalibraci Do asn ukon kalibraci Je umo n n p stup k ostatn m nab dk m Po...

Page 143: ...te ve ker osobn ochrann pom cky vy adovan pro manipulaci s p slu n mi chemik liemi Viz aktu ln datov bezpe nostn listy MSDS SDS U P O Z O R N N P ed vyjmut m elektrody v dy odstra te plnic z tku elekt...

Page 144: ...knout tmav povlak bromidu st brn ho AgBr Tento povlak neovliv uje v sledky m en pokud nen takto kontaminov no v ce ne 90 povrchu P i v m n elektrolytu a membr ny st brnou anodu vizu ln kontrolujte Pok...

Page 145: ...ezera mezi membr nou a katodou Ut hn te p dr nou matici elektrody Membr na je nespr vn nainstalov na a vzniklo riziko zne i t n elektrolytu Vym te elektrolyt a do t la sn ma e znovu nainstalujte membr...

Page 146: ...ebovan Vym te membr nu Na st brn trubici je tmav zelen usazenina bromidu st brn ho AgBr Vyle t te trubici m kk m brusn m prost edkem 400 a 600 a vym te membr nu V znamn nestabilita v re imu m en V ko...

Page 147: ...tak aby odpov dal specifikac m Na katod jsou usazeniny nebo ne istoty Katodu vy ist te m kk m a sav m had kem Vym te membr nu Proud sn ma e je b hem m en nulov V sn ma i nen elektrolyt net snost sn m...

Page 148: ...V stra n ikony se objevuj na prav stran hlavn ho displeje pod jednotkou m en Varov n neovlivn provoz nab dek vys l n a v stup Chcete li varov n zobrazit stiskn te tla tko menu a zvolte DIAGNOSTICS Dia...

Page 149: ...a p slu enstv naleznete v sti t kaj c se n hradn ch d l a p slu enstv dokumentace ke kontrol ru Pozn mka sla produkt a polo ek se mohou v n kter ch prodejn ch oblastech li it Obra te se na p slu n ho...

Page 150: ...geving vooraf worden gewijzigd Sensor Specificatie Gegevens Sensor Noryl Sensorbehuizing met optionele onderdompeling Roestvrij staal 316L Afmetingen 144 x 144 x 181 mm 5 7 x 5 7 x 7 1 inch Gewicht 1...

Page 151: ...eninformatie G E V A A R Geeft een potentieel gevaarlijke of dreigende situatie aan die als deze niet kan worden voorkomen kan resulteren in dodelijk of ernstig letsel W A A R S C H U W I N G Geeft ee...

Page 152: ...n voedingwaterboilers condensatoren en thermische apparatuur in het algemeen waarbij water gebruikt wordt als vloeistof voor warmteoverdracht Productcomponenten Controleer of alle componenten zijn ont...

Page 153: ...r of er zich geen onzuiverheden of luchtbellen in het elektrolyt bevinden 3 Duw de elektrode voorzichtig in de sensorbehuizing tot er weerstand is Forceer de elektrode niet 4 Breng de borgmoer aan en...

Page 154: ...ingangsconnector 4 Sluit de monsteruitgang aan op de monster uitgangsconnector Afbeelding 2 De sensor monteren 1 Connector sensorkabel 5 Doorstroomkamer 2 Klemmoer sensor 6 1 4 NPT connector monsteri...

Page 155: ...de barri re op zijn plaats blijven L E T O P Potenti le schade aan apparaat Delicate interne elektronische componenten kunnen door statische elektriciteit beschadigd raken wat een negatieve invloed o...

Page 156: ...156 Nederlands...

Page 157: ...teren en druk op enter Houd de pijltoetsen omhoog of omlaag ingedrukt om cijfers tekens of interpunctie in te voeren Druk op de pijltoets naar rechts om naar de volgende ruimte te gaan Optie Beschrijv...

Page 158: ...ibratie en druk op enter 5 De bruto temperatuurwaarde wordt weergegeven Druk op enter 6 Voer de juiste waarde in indien deze afwijkt van hetgeen wordt weergegeven en druk op enter 7 Druk op enter om d...

Page 159: ...oegangscode is ingeschakeld in menu security beveiliging voor de controller voer dan de toegangscode in 4 Selecteer AIR CAL luchtkalibratie en druk op enter 5 Selecteer de optie voor het uitgangssigna...

Page 160: ...arde te wijzigen 5 Druk met de sensor in het procesmonster op enter De gemeten waarde wordt weergegeven Wacht tot de waarde is gestabiliseerd en druk op enter 6 Meet met een gecertificeerd tweede veri...

Page 161: ...BRATE sensorinstellingen kalibreren 2 Als de toegangscode is ingeschakeld in menu security beveiliging voor de controller voer dan de toegangscode in 3 Selecteer KAL OPTIES en druk op enter 4 Gebruik...

Page 162: ...ebruik geen versleten membraan Verwijder de sensor 1 Schakel de monstertoevoer uit 2 Schroef de sensorkabelconnector van de sensor los Draai niet aan de kabel of de sensor zelf 3 Schroef de klemmoer v...

Page 163: ...embraan moet de zilveren anode visueel gecontroleerd worden Als meer dan 2 3 van het oppervlak bedekt is met zilverbromide moet de elektrode verjongd worden Polijst hiervoor de met afzetting bedekte g...

Page 164: ...aan is niet juist geplaatst wat een risico op vervuiling van het elektrolyt vormt Vervang het elektrolyt en plaats het membraan opnieuw vingervast zo ver als het gaat op het sensorlichaam Onvoldoende...

Page 165: ...eren slang bevat een donkergroene afzetting van zilverbromide AgBr Polijst de slang met een zacht schuurmiddel N 400 tot 600 en vervang het membraan Belangrijke instabiliteit in de meetstand De kabelc...

Page 166: ...het binnen de specificaties valt Er zit aanslag of er zitten deeltjes op de kathode Reinig de kathode met een zachte absorberende doek Vervang het membraan De sensorstroom is nul tijdens een meting E...

Page 167: ...waarschuwingen Een waarschuwing heeft geen effect op de werking van menu s relais en uitgangen Druk op toets menu en selecteer DIAGNOSTICS diagnostiek om waarschuwingen te bekijken Selecteer hierna e...

Page 168: ...pleeg het gedeelte over reserveonderdelen en accessoires van de documentatie bij de controller voor onderdelen en accessoires van de controller Opmerking Product en artikelnummers kunnen in bepaalde v...

Page 169: ...Specifikation Detaljer Sensorhoveddel Noryl Sensorhus med option til neds nkning Rustfrit st l 316L Dimensioner 144 x 144 x 181 mm 5 7 x 5 7 x 7 1 V gt 1 7 kg 3 75 lb Elektrode Katode guld Anode s lv...

Page 170: ...p nogen anden m de end den der er angivet i denne manual Brug af sikkerhedsoplysninger F A R E Angiver en eventuel eller overh ngende farlig situation der vil medf re d dsfald eller alvorlige kv stels...

Page 171: ...orvarmere kondensatorer og generelt alt termisk udstyr der anvender vand som varmeoverf rselsv ske Produktkomponenter S rg for at alle komponenter er modtaget Kontakt producenten eller en forhandler m...

Page 172: ...es modstand Brug ikke kraft 4 Mont r og sp nd holdem trikken med h ndkraft 5 Mont r pakningen og p fyldningsskruen for at forhindre elektrolytl kage eller udefrakommende forurening fra pr ven Sp nd fo...

Page 173: ...r Forbind sensorstikket til kontrolleren A D V A R S E L Risiko for livsfarligt elektrisk st d S rg altid for at sl str mmen til instrumentet fra n r du tilslutter str mf rende elementer A D V A R S E...

Page 174: ...1 S rg for at tilslutte alle jordforbundne sensorsk rmkabler til de jordforbundne afsk rmningsskruer p controllerens kabinet BEM RK Hvis sensorkablet ikke er langt nok til at n controlleren er et sam...

Page 175: ...rene for l kager og reparer eventuelle l kager 4 T nd for str mmen til kontrolenheen 5 Foretag de nskede valg i menuerne n r controlleren starter Brugernavigering Se styringsdokumentationen for beskri...

Page 176: ...t for ger tiden for sensorsignalets reaktion p faktiske ndringer i processen LOG SETUP LOGOPS TNING Indstiller tidsintervallet for datalagring i dataloggen 5 30 sekunder 1 2 5 10 15 standard 30 60 min...

Page 177: ...NS Kalibreringsfunktioner skal der indtastes et bruger id Se ndre kalibreringsfunktioner p side 179 11 V lg p sk rmbilledet NEW SENSOR Ny sensor om sensoren er ny Funktion Beskrivelse JA Sensoren har...

Page 178: ...udgangstilstand er indstillet p pause eller til at overf re skal man v lge forsinkelsestiden n r udgangene vender tilbage til den aktive tilstand Kalibrering med procespr ven Sensoren kan forblive i...

Page 179: ...tilbage til kalibreringen G UD AF KAL Afslut kalibreringen midlertidigt Der er mulighed for adgang til andre menuer Tryk p tasten menu og v lg SENSOR SETUP Ops tning af sensor for at vende tilbage til...

Page 180: ...vendigt for at beskytte dig mod de kemikalier du bruger Se de aktuelle sikkerhedsdataark MSDS SDS for sikkerhedsprotokoller B E M R K N I N G Fjern altid elektrolytp fyldningsproppen inden afmontering...

Page 181: ...s osv kan en m rk bel gning af s lvbromid AgBr d kke en del af s lvanoden Bel gningen p virker ikke m lingen medmindre mere end 90 af overfladen er kontamineret Efterse s lvanoden ved udskiftning af e...

Page 182: ...rt mellemrum mellem membranen og katoden Sp nd elektrodens holdem trik Membranen er monteret forkert og medf rer en risiko for elektrolytforurening Udskift elektrolytten og genmont r membranen p senso...

Page 183: ...ntrolleres Membranen er slidt Udskift membranen Der er en m rkegr n bel gning af s lvbromid AgBr p s lvr ret Pol r r ret med fint sandpapir korn 400 til 600 og udskift membranen Betydelig ustabilitet...

Page 184: ...r s rg for at pr ven er inden for specifikationen Der er slam eller partikler p katoden Reng r katoden med en bl d og absorberende serviet Udskift membranen Sensorstr mmen er null under m ling Der er...

Page 185: ...p virker ikke driften af menuer rel og stikkontakter Tryk p tasten menu og v lg DIAGNOSTICS Diagnosticering for at f vist fejl V lg derefter enheden for at f vist eventuelle problemer p den p g ldende...

Page 186: ...reservedele og tilbeh r til controlleren BEM RK Produkt og varenumre kan variere i visse salgsregioner Kontakt den relevante distribut r eller se virksomhedens webside for kontaktinformation Reservede...

Page 187: ...elande Sensor Specifikation Tekniska data Givarkropp Noryl Givarkropp med valfri neds nkning Rostfritt st l 316L M tt 144 x 144 x 181 mm 5 7 x 5 7 x 7 1 tum Vikt 1 7 kg 3 75 lb Elektrod Katod guld ano...

Page 188: ...om specificeras i den h r handboken Anm rkning till information om risker F A R A Indikerar en potentiellt eller verh ngande riskfylld situation som kommer att leda till livsfarliga eller allvarliga s...

Page 189: ...dast anv ndas i industriella milj er Produkt versikt Den h r enkanalsanalysatorn anv nds f r m tning av l st syre i matarvatten till ngpannor f rv rmare kondensorer och i allm nhet all v rmeutrustning...

Page 190: ...f rsiktigt elektroden in i givarkroppen tills du k nner ett motst nd Tvinga inte in elektroden 4 Installera och dra t l smuttern f r hand 5 Installera packningen och fyllningsskruven f r att f rhindra...

Page 191: ...nloppsanslutning medf ljer ej 3 Monterad givare 7 M4 skruvar medf ljer ej 4 Packning 8 1 4 in NPT provutloppsanslutning medf ljer ej Installera styrenheten Se dokumentationen till styrenheten f r anvi...

Page 192: ...s av statisk elektricitet vilket kan leda till f rs mrad funktion hos instrumentet eller till att det inte fungerar Proceduren illustreras i figurerna nedan och Tabell 1 Anslut alla givarens jord sk r...

Page 193: ...Svenska 193...

Page 194: ...ange siffror tecken eller skiljetecken Flytta till n sta position genom att trycka p h ger pilknapp Alternativ Beskrivning EDIT NAME Redigera namn ndrar namnet som motsvaras av givaren l ngst upp p d...

Page 195: ...a nollpunkt 1 Ta bort givaren fr n processen och sk lj den i destillerat vatten 2 Tryck p knappen menu meny och v lj SENSOR SETUP CALIBRATE givarinst llning kalibrera 3 Om l senordet r aktiverat i sty...

Page 196: ...strumentets anv ndarhandbok 6 Anv nd en precisionscertifierad barometer och m t atmosf rstrycket d r analysatorn r placerad Anv nd pilknapparna f r att ange det h r v rdet om det skiljer sig fr n det...

Page 197: ...7 Granska kalibreringsresultatet PASS godk nt givaren r kalibrerad och kalibreringsfaktorn visas FAIL underk nt kalibreringen ligger utanf r de godk nda gr nserna Reng r givaren och f rs k igen Mer i...

Page 198: ...a kalibrering i dagar m nader eller r v lj nskat intervall i listan ANV ID VID KALIBR L gger ett anv ndar ID till kalibreringsdata Yes ja eller No nej standardinst llning ID t anges under kalibreringe...

Page 199: ...3 Ta bort givaren Byta ut membranet Se stegen som f ljer och Figur 4 f r byte av membran 1 Skruva loss fyllningsskruven och ta bort den tillsammans med packningen 2 Skruva loss l smuttern 3 Ta f rsik...

Page 200: ...yts ut ska du kontrollera silveranoden Om mer n 2 3 av ytan r t ckt med silverbromid s beh ver elektroden f rnyas F r att g ra det ska du f rsiktigt polera omr dena som r t ckta med avlagringen med et...

Page 201: ...membranet och katoden r f r stort Dra t l smuttern till elektroden Membranet r felaktigt installerat vilket g r att det finns en risk f r att elektrolyten f rorenas Byt ut elektrolyten och installera...

Page 202: ...ranet Det finns en m rkgr n avlagring av silverbromid AgBr p silverr ret Polera r ret med ett mjukt sandpapper nr 400 till 600 och byt ut membranet Betydande instabilitet i m tl get Det finns vatten e...

Page 203: ...r ndra provet s att det ligger inom specifikationerna Slam eller partiklar p katoden Reng r katoden med en mjuk och absorberande torkduk Byt ut membranet Givarens str m r noll under m tningen Det finn...

Page 204: ...utg ngarnas funktion Visa varningar genom att trycka p knappen menu meny och v lj Diagnostics instrumentdiagnostik V lj sedan enheten f r att se eventuella fel som f rknippas med den enheten Varnings...

Page 205: ...e eller se f retagets webbsida f r att f kontaktinformation Utbytesdelar och tillbeh r Beskrivning Produktnr Kartong med 4 f rmonterade membran 09185 A 3500 Kalibreringshylsa 09182 A 1200 Fyllningsskr...

Page 206: ...4 x 181 mm 5 7 x 5 7 x 7 1 in Paino 1 7 kg 3 75 lb Elektrodi Katodi kultaa Anodi hopeaa Kalvo PFA Kalvon pidin Noryl N ytteen virtausnopeus 4 10 litraa tunti N ytteen l mp tila 20 60 C 4 120 F N yttee...

Page 207: ...oka aiheuttaa kuoleman tai vakavan vamman V A R O I T U S Osoittaa potentiaalisesti tai uhkaavasti vaarallisen tilanteen joka jos sit ei v ltet voisi johtaa kuolemaan tai vakavaan vammaan V A R O T O...

Page 208: ...ikista l mmityslaitteista jotka k ytt v t vett l mm nsiirtonesteen Tuotteen osat Varmista ett laitteen mukana on toimitettu kaikki tarvittavat osat Jos joku tarvikkeista puuttuu tai on viallinen ota v...

Page 209: ...akota elektrodia 4 Asenna ja kirist pid kemutteri k sin 5 Asenna tiiviste ja t ytt ruuvi jotta elektrolyytti ei p se vuotamaan eik se kontaminoidu n ytteen vaikutuksesta Kirist ruuvi huolellisesti ruu...

Page 210: ...k ytt ohjeista ohjeita kiinnitykseen ja johdotukseen Vahvistimen liitt minen anturiin V A R O I T U S S hk iskun vaara Kytke aina laite irti verkkovirrasta ennen s hk liit nt jen tekemist V A R O I T...

Page 211: ...o Taulukko 1 Varmista ett liit t kaikki anturin maadoitus suojajohtimet vahvistimen kotelon maadoitusruuveihin Huomautus Jos anturikaapeli ei ole niin pitk ett se ulottuisi vahvistimeen tarvitaan lis...

Page 212: ...m ritt virtausnopeuden 3 Tutki letkujen kytkent vuotojen varalta ja tuki l ytyneet vuodot 4 Johda vahvistimeen virta 5 Tee tarvittavat valikkovalinnat vahvistimen k ynnistyess Navigoiminen N pp imist...

Page 213: ...ikav lilt v lilt 0 ei vaikutusta oletusarvo ja 60 sekuntia signaaliarvon keskiarvo 60 sekunnin aikana Suodatin pident aikaa joka vaaditaan siihen ett anturin signaali reagoi muutoksiin prosessissa LOG...

Page 214: ...a yrit uudelleen uudella viiteratkaisulla Lis tietoa saa kohdasta Vianm ritys sivulla 219 9 Jos kalibrointi onnistui jatka painamalla enter n pp int 10 Jos k ytt j tunnuksen valinnaksi on asetettu YES...

Page 215: ...t nollataan NO EI Anturi on kalibroitu aikaisemmin t ll ohjaimella 13 Palauta anturi prosessiin ja paina enter n pp int L ht signaali palaa aktiiviseen tilaan ja mittausn yt ss esitet n mitattu n yte...

Page 216: ...on aloitettava alusta BACK TO CAL TAKAISIN KALIBROINTIIN Palaa kalibrointiin LEAVE CAL POISTU KALIBROINNISTA Poistu kalibroinnista tilap isesti Muut valikot ovat k ytett viss Palataksesi kalibrointiin...

Page 217: ...tuvia suojavarusteita Lue turvallisuusprotokollat ajan tasalla olevista k ytt turvatiedotteista MSDS SDS H U O M A U T U S Irrota aina elektrolyytin t ytt tulppa ennen elektrodin irrottamista l koske...

Page 218: ...ektrodin hopea anodista saattaa olla tumman hopeabromidikerroksen AgBr peitossa T m ei vaikuta mittaukseen ellei yli 90 pinnasta ole kerroksen peitossa Kun vaihdat elektrolyytti ja kalvoa tarkista hop...

Page 219: ...ja ulkoa ja asenna se sormivoimin Kultapinnoite on naarmuuntunut tai vaurioitunut Vaihda elektrodi Katodissa sakkaa tai partikkeleita Puhdista katodi pehme ll ja imukykyisell liinalla Huuhtele kalvo K...

Page 220: ...da elektrolyytti ja kalvo H iri t yleens kyseess on H2S Jos H2S n tai muun likaavan aineen taso on vakaa ota sen pitoisuus huomioon m ritelless si saturoituneen O2 n pitoisuus Virhe kalibroinnissa tai...

Page 221: ...ista RESET CAL HISTORY NOLLAA KALIBROINTIHISTORIA Vain huoltok ytt n Nollaa anturin kalibrointihistorian Kaikki aikaisemmat kalibrointitiedot menetet n SENSOR SIGNALS SENSORIN SIGNAALIT Esitt lukeman...

Page 222: ...ria ei ole kalibroitu Kalibroi anturi CAL IN PROGRESS KALIBROINTI K YNNISS Kalibrointi on k ynnistynyt mutta ei p ttynyt Palaa kalibrointiin Varaosat ja varusteet Ohjaimen osat ja varusteet on esitelt...

Page 223: ...i adatok R szletes adatok rz kel test Noryl rz kel test opcion lis bemer l ssel Rozsdamentes ac l 316L M retek 5 7 h velyk 144 mm x 5 7 h velyk 144 mm x 7 1 h velyk 181 mm T meg 1 7 kg 3 75 font Elekt...

Page 224: ...vetlen l vesz lyes helyzetet jelez amely hal lhoz vagy s lyos s r l shez vezet F I G Y E L M E Z T E T S Lehets ges vagy k zvetlen l vesz lyes helyzetet jelez amely hal lhoz vagy s lyos s r l shez vez...

Page 225: ...aszn lj k A term k r szegys gei Gy z dj n meg arr l hogy minden r szegys get megkapott Ha valamelyik t tel hi nyzik vagy s r lt forduljon azonnal a gy rt hoz vagy a forgalmaz hoz Rendszerelemek Vez rl...

Page 226: ...bor k az elektrolitban 3 vatosan nyomja az elektr d t az rz kel test be am g ellen ll st nem rez Ne er ltesse az elektr d t 4 K zzel szerelje fel s szor tsa meg a tart any t 5 Szerelje fel a t m t st...

Page 227: ...a bemeneti csatlakoz j hoz 4 Csatlakoztassa a minta kimenet t a minta kimeneti csatlakoz j hoz 2 bra Az rz kel felszerel se 1 rz kel k bel csatlakoz ja 5 tfoly cella 2 rz kel t r gz t csavaranya 6 1 4...

Page 228: ...Z S Lehets ges k rosod s a k sz l kben Az rz keny bels elektronikus rendszerelemek megs r lhetnek a statikus elektromoss g k vetkezt ben amely cs kkent m k d k pess get vagy esetleges le ll st eredm n...

Page 229: ...Magyar 229...

Page 230: ...ombok haszn lat val v lasszon ki egy opci t majd nyomja meg az Enter gombot Sz mok karakterek vagy r sjelek be r s hoz nyomja meg a fel vagy a le ny lgombot Nyomja le a jobbra ny lgombot hogy a k vetk...

Page 231: ...yes rt ket ha az k l nb zik a megjelent rt kt l majd nyomja meg az Enter gombot 7 A kalibr l s j v hagy s hoz nyomja meg az Enter gombot Megjelenik a h m rs klet elt r s rt ke Nullapont kalibr l sa Az...

Page 232: ...a LEVEG KALIB men pontot s nyomja meg az Entergombot 5 A kalibr l s sor n v lassza ki az opci t a kimeneti jelhez Opci Le r s ACTIVE AKT V A k sz l k a kalibr l si elj r s sor n elk ldi a jelenleg m...

Page 233: ...an nyomja meg az Enter gombot A m rt rt k jelenik meg V rja meg am g az rt k stabiliz l dik majd nyomja meg az Enter gombot 6 Hiteles tett m sodlagos ellen rz k sz l kkel m rje meg a minta koncentr ci...

Page 234: ...4 A ny lgombok haszn lat val v lasszon ki egy opci t majd nyomja meg az Enter gombot Opci Le r s KALIB EML KEZT Eml keztet be ll t sa a k vetkez kalibr l shoz nap h nap vagy v a list r l v lassza ki...

Page 235: ...rz kel t r gz t csavarany t az raml si kamr r l s t vol tsa el az rz kel t 3 bra 3 bra Szerelje le az rz kel t Cser lje ki a membr nt Az rz kel membr nj nak cser j hez l sd az al bbi l p seket s a k...

Page 236: ...zze az ez st an dot Ha az elektr da fel let nek t bb mint 2 3 r sz t ez stbromid bevonat bor tja az elektr d t fel kell j tani Ennek elv gz s hez puha csiszol anyaggal 400 as 600 as finomat csiszolja...

Page 237: ...s a kat d k z tt Szor tsa meg az elektr d t tart any t A membr n nem j l van beszerelve ami az elektrolit szennyez d s t okozhatja Cser lje ki az elektrolitot s szerelje be jra a membr nt az rz kel te...

Page 238: ...omid AgBr bevonat l that az ez st cs v n Csiszolja meg a cs vet puha csiszol anyaggal 400 as 600 as s helyezze vissza a membr nt Fontos instabilit s m r si m dban Az rz kel k bel csatlakoz j ban v z v...

Page 239: ...y m dos tsa a mint t gy hogy a specifik ci ban megadott rt keken bel l maradjon Iszap vagy apr r szecsk k tal lhat k a kat don Tiszt tsa meg a kat dot puha felsz v sz vettel Cser lje ki a membr nt A m...

Page 240: ...oldal n a m r si rt k alatt jelennek meg A figyelmezet sek nem befoly solj k a men k rel k illetve a kimenetek m k d s t A figyelmeztet sek megtekint s hez nyomja meg a Men gombot majd v lassza a DIAG...

Page 241: ...kaival kapcsolatos tudnival kat l sd a vez rl dokument ci j nak Cserealkatr szek s tartoz kok c m fejezet ben Megjegyz s A term k s cikksz mok rt kes t si r gi nk nt elt rhetnek L pjen kapcsolatba a m...

Page 242: ...49 243 254 244 256 Startup 249 260 249 316L 144 x 144 x 181 5 7 x 5 7 x 7 1 1 7 3 75 4 10 20 60 C 4 120 F 0 90 EN 61326 1 2006 EN 61010 1 2010 0 2 ppm 0 5 ppb 0 5 ppb 2 1 ppb 1 40 ppb 30 0 45 C 32 113...

Page 243: ...243...

Page 244: ...EMC EMC A A 244...

Page 245: ...MSDS SDS 1 1 2 5 3 4 5 1 245...

Page 246: ...2 1 2 3 4 2 1 5 2 6 1 4 NPT 3 7 4 4 8 1 4 NPT 246...

Page 247: ...1 1 1 2 3 4 5 6 7 8 9 10 11 12 247...

Page 248: ...248...

Page 249: ...Startup 242 1 2 3 4 5 CONFIGURE 1 menu SENSOR SETUP CONFIGURE 2 EDIT NAME 16 SENSOR S N 16 MEAS UNITS PRESSURE UNITS TEMP UNITS C F FILTER 0 60 60 249...

Page 250: ...TUP 5 30 1 2 5 10 15 30 60 RESET DEFAULTS 1 2 menu SENSOR SETUP CALIBRATE 3 4 1 PT TEMP CAL enter 5 enter 6 enter 7 enter 1 2 menu SENSOR SETUP CALIBRATE 3 4 ZERO CAL enter 5 ACTIVE HOLD TRANSFER 6 en...

Page 251: ...7 enter 8 PASS FAIL 256 9 enter 10 ID YES CAL OPTIONS ID 253 11 NEW SENSOR YES NO 12 enter 1 2 menu SENSOR SETUP CALIBRATE 3 4 AIR CAL enter 5 ACTIVE HOLD TRANSFER 6 enter 7 enter 8 enter 9 PASS 251...

Page 252: ...enter 11 ID YES CAL OPTIONS ID 12 NEW SENSOR YES NO 13 enter 1 menu SENSOR SETUP CALIBRATE 2 3 SAMPLE CAL enter 4 ACTIVE HOLD TRANSFER 5 enter enter 6 enter 7 PASS FAIL 256 8 enter 9 ID YES CAL OPTION...

Page 253: ...O 11 enter back 1 back QUIT CAL BACK TO CAL LEAVE CAL menu SENSOR SETUP 2 enter ID 1 menu SENSOR SETUP CALIBRATE 2 3 CAL OPTIONS enter 4 CAL REMINDER OP ID ON CAL ID YES NO 9582 2 2 5 menu SENSOR SETU...

Page 254: ...MSDS SDS 1 2 3 3 3 254...

Page 255: ...4 1 2 3 4 5 6 245 4 3 12 AgBr 90 2 3 N 400 600 30 255...

Page 256: ...6 C 43 F 35 C 95 F 260 256...

Page 257: ...AgBr N 400 600 H2S 4 257...

Page 258: ...H2S H2S O2 4 AgBr 400 600 0 10 000 2 258...

Page 259: ...2 2 menu DIAGNOSTICS 3 3 40 0 TEMP TOO HIGH T 50 C TEMP TOO LOW T 0 C CURRENT TOO HIGH 200 CURRENT TOO LOW 0 5 A 259...

Page 260: ...3 CAL OVERDUE REPLACE SENSOR 365 DIAG TEST NOT CALIBRATED CAL IN PROGRESS 4 09185 A 3500 09182 A 1200 09078 C 1020 09078 C 1030 09182 A 1000 09078 C 1010 25 09181 A 3600 09078 A 2000 460150 21951 260...

Page 261: ...Sens r Teknik zellik Ayr nt lar Sens r g vdesi Noryl Opsiyonel dald rmal sens r g vdesi Paslanmaz elik 316 L Boyutlar 144 x 144 x 181 mm 5 7 x 5 7 x 7 1 in A rl k 1 7 kg 3 75 lb Elektrot Katot alt n A...

Page 262: ...n ba ka bir ekilde kullanmay n veya kurmay n Tehlikeyle ilgili bilgilerin kullan lmas T E H L K E Olmas muhtemel veya yak n bir zamanda olmas ndan korkulan engellenmedi i takdirde l m veya ciddi yaral...

Page 263: ...kazan besleme suyundaki ekonomiz rdeki kondens rdeki ve genel olarak suyu s transfer s v s bi iminde kullanan t m termal ekipmanlardaki z nm oksijenin l m nde kullan l r r n bile enleri B t n bile enl...

Page 264: ...f e itin Elektrotu zorlamay n 4 Tutma somununu elle tak n ve s k n 5 Numuneyi elektrik s z nt s veya harici kirlenmeden korumak i in conta ve doldurma vidas n tak n Viday bir tornavidayla dikkatli bir...

Page 265: ...jlanm sens r 7 M4 vidalar temin edilmedi 4 Conta 8 1 4 in NPT numune k konekt r temin edilmedi Kontrol r n tak lmas Monte etme ve kablolama talimatlar i in kontrol r belgesine bak n Sens r kontrol re...

Page 266: ...ik elektrikten zarar g rebilir ve bu durum cihaz performans n n d mesine ya da cihaz n ar zalanmas na neden olabilir A a da g sterilen resimli ad mlar izleyin ve uraya bak n Tablo 1 T m sens r toprak...

Page 267: ...T rk e 267...

Page 268: ...rlemek i in sa y n tu una bas n Se enek A klama ADI D ZENLE l m ekran n n zerindeki sens re kar l k gelen ad de i tirir Ad harflerin say lar n bo luklar n veya noktalama i aretlerinin herhangi bir kom...

Page 269: ...ni se in 3 Parola kontrol r n g venlik men s nde etkinle tirilmi se parolay girin 4 SIFIR KAL se ene ini se in ve enter tu una bas n 5 Kalibrasyon s ras nda k sinyali se ene ini belirleyin Se enek A k...

Page 270: ...kalibrasyon kapa na dikey olarak membran a a do ru gelecek bi imde yerle tirin Kalibrasyon kapa n s k n ve enter tu una bas n 8 De erin sabitlenmesini bekleyip enter tu una bas n 9 Kalibrasyon sonucu...

Page 271: ...de i tirilmesi sayfa 271 10 YEN SENS R ekran nda sens r n yeni olup olmad n se in Se enek A klama EVET Sens r bu kontrol rle daha nce kalibre edilmedi Sens rler i in al ma g nleri ve nceki kalibrasyo...

Page 272: ...ve l m al nmaz Bekleme modundan kmak i in men tu una bas n ve SENS R AYARLARI BEKLEMEDEN IK elerini se in Bak m U Y A R I Birden fazla tehlike Belgenin bu b l m nde a klanan g revleri yaln zca yetkili...

Page 273: ...uraya bak n ekil 4 1 Doldurma vidas n z n ve contayla birlikte kar n 2 Tutma somununu kar n 3 Elektrotu sens r g vdesinden dikkatlice kar n 4 Sens r g vdesinde kalan t m elektroliti bo alt n 5 Eski me...

Page 274: ...m anotun y zeyini kontrol edin Y zeyin 2 3 nden fazlas g m brom rle kapl ysa elektrot temizlemesi yapmak gerekir Bunu yapmak i in tortuyla kaplanm b lgeleri yumu ak bir z mparayla N 400 to 600 hafif e...

Page 275: ...lde tak n Alt n y zey izilmi veya zarar g rm Elektrotu de i tirin Katotta amur veya par ac klar var Katotu yumu ak ve emici bir kuma la silin Membran durulay n Sens r kar l rken kablo ve ba lant lar z...

Page 276: ...durma vidas kontrol edin membran ve elektroliti de i tirin Elektrolit s z nt s Vidalama b lgelerini membran doldurma vidas kontrol edin membran ve elektroliti de i tirin Etkile imler esas olarak H2S i...

Page 277: ...ni se in Tablo 2 Sensor DIAG TEST Sens r TANILAMA TEST men s Se enek A klama MODULE INFORMATION MOD L B LG LER Sens r mod l n n s r m n ve seri numaras n g sterir SENSOR INFORMATION SENS R B LG LER Ku...

Page 278: ...de i tirip sens r kalibre edin Kalibrasyon sonucu ba ar l ise TANI TEST men s nde membran g n say s n s f rlay n KAL BRE ED LMED Sens r kalibre edilmedi Sens r kalibre edin KAL S R YOR Bir kalibrasyo...

Page 279: ......

Page 280: ...orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route d...

Reviews: