Option
Description
Hold during calibration
On by default, this stops any output from the instrument during the calibration
process to avoid sending invalid information to any connected device.
Automatic calibration
stop
If selected, when the stability criteria is reached, the calibration process stops
automatically.
2.
Press
OK
to start calibration
• A calibration screen will be displayed showing current measurement data which is continually
refreshed.
• The value “% ideal current” is a percentage of the current against the ideal current for the
membrane type selected. If this percentage is not within the accepted range, an error message
is displayed and the calibration process fails. A warning message can be displayed when this
value is close to the boundaries, but when calibration can be accepted.
• The message is first displayed in the result box. The dialog box with the error message or the
warning is displayed when the finish button is pressed.
• The value “% last calibration” shows the ratio between the current measurement and the
previous sensor calibration.
• The value “% variation” indicates the variation during the last 3 measurements, which is the
stability of the measurements. A variation as low as possible is needed for a precise
calibration.
• The display shows the actual calibration parameters, and the actual readings (temperature,
barometer, current).
Barometric pressure calibration
Note: The barometric pressure sensor has been factory calibrated but should be periodically verified with a
precision certified barometer. This is only necessary if measuring in gas phase with fraction units (%, ppm).
The upper box shows the barometric pressure as measured by the instrument.
Using a precision certified barometer measure barometric pressure in the location where the
measuring instrument is used. Compare the values and if they are the same press
Cancel
, otherwise
enter the new barometric pressure value in the lower box and press
Validation
to validate the new
setting.
Services menu
There are a number of options available from this menu with the key options detailed below.
Calibration timer
The instrument can automatically remind the user when the next sensor calibration is due.
• To enable the timer, select
Enable
and enter a delay in days.
• The display shows the current instrument date and time, next calibration due date and time, and
the remaining days.
The next calibration date is updated when the sensor is calibrated.
Service timer
The sensor attached to your instrument will require periodic servicing and maintenance. The
instrument can automatically remind the user when the next sensor service is due.
• To enable the timer, select
Enable
and enter a delay in days.
• The display shows the current instrument date and time, the next sensor service due date and
time, and the remaining days.
After a service press the
Service done
button to update the next service date.
22
English
Summary of Contents for ORBISPHERE 410
Page 179: ...ESD 2000 m 6562 ft 2000 m 179...
Page 180: ...GFCI GFI PE 100 240 VAC 5 VDC PE PE 1 U 2 3 2 4 1 U 2 3 4 2 5 180...
Page 181: ...1 3 4 5 6 7 1 2 3 6 4 4 5 4 1 4 6 7 1 1 5 USB A 2 USB B 4 6 2 3 Ethernet 7 1 4 8 181...
Page 182: ...NEMA IP NEMA 4X IP66 1 Ethernet 2 25 mm 8 mm 3 4 1 4 7 2 5 8 3 6 5 6 10 30 VDC 8 BINDER 182...
Page 184: ...1 4 2 3 1 2 3 3 1 2 3 4 1 2 3 4 3 4 5 4 2 6 1 7 ORBISPHERE J8 P8 4 185 J7 J8 6 186 7 186 184...
Page 188: ...32605 ORBISPHERE TC LED 10 1 menu Up Main Close Help Enter 188...
Page 190: ...1 Configure View Conf Channel temperature Grid y z Auto scale Clean 2 190...
Page 191: ...1 Config instrument Measurement Continuous 1 EC EC ppm ppb 5 0 1 2 3 HOT 90 2 TC TC EC TC 191...
Page 196: ...2 OK ideal current 3 O2 O2 30 O2 O2 O2 O3 196...
Page 197: ...O3 O2 O2 O3 O2 O2 O3 O3 O3 TC 1 Modify Modify 2 OK ideal current 3 197...
Page 198: ...ppm Cancel Validation Enable Enable Service done RS485 PROFIBUS DP USB HTTP TCP IP 198...
Page 199: ...Hach 199...
Page 224: ...2000 6562 2000 224...
Page 225: ...GFCI GFI PE 100 240 5 1 U 2 3 4 225...
Page 226: ...1 U 2 3 4 5 1 3 4 5 6 7 1 2 3 6 4 4 5 4 1 4 6 7 226...
Page 227: ...1 1 5 USB A 2 4 USB B 6 2 3 Ethernet 7 1 4 8 NEMA IP NEMA 4X IP66 1 Ethernet 2 25 8 3 4 227...
Page 228: ...1 4 7 2 5 8 3 6 5 6 10 30 8 BINDER 2 BINDER 1 10 30 2 3 4 5 6 10 30 7 10 30 8 3 228...
Page 231: ...Ethernet RJ 45 ethernet ethernet 1 227 P3 4 6 7 J8 6 7 8 J7 9 J8 J7 16 J7 8 231...
Page 233: ...J4 1 J5 2 J6 3 40 ORBISPHERE 29089 ORBISPHERE 32605 ORBISPHERE 233...
Page 234: ...10 Menu Up Main Close Help CAP Enter Security 234...
Page 244: ...last calibration variation 3 ppm Cancel Validation Services Enable Enable Service done 244...
Page 356: ......
Page 357: ......