![Hach ORBISPHERE 410 Basic User Manual Download Page 13](http://html1.mh-extra.com/html/hach/orbisphere-410/orbisphere-410_basic-user-manual_3155471013.webp)
TC sensors
For TC sensor installation, servicing and maintenance ensure you follow the instructions in the
TC
Sensor Installation and Maintenance
manual that was supplied with the instrument. Pay particular
attention to the installation and connection of the purge gas supply.
N O T I C E
Do not place the TC sensor into a liquid sample until a constant supply of dry purge gas has been connected, as
liquid could condense inside the measuring chamber and cause damage to the thermal conductor chip.
To ensure the continuation of purge gas while the sensor is in contact with the sample, it is highly
recommended to use a backup purge gas cylinder with an automatic changeover valve that activates
when the first cylinder is empty.
The use of an ORBISPHERE Model 29089 gas regulator (or similar) is also recommended to deliver
a constant, pressure regulated supply of dry purge gas to the sensor, filtered to 40 μm.
In addition, and to prevent any damage to the sensor electronics, the use of a purge safety backup
unit (ORBISPHERE Model 32605) is highly recommended to ensure the supply of purge gas remains
uninterrupted to the sensor in the event of a mains power outage.
The above ORBISPHERE accessories are explained in more detail in the
TC Sensor Installation
and Maintenance
manual.
User interface
Instrument controls
The instrument front panel provides:
• A touch screen acting as display, touch pad and keyboard.
• A LED, showing when the instrument is on.
Turning instrument On and Off
There is no power switch on the instrument. The mains must be disconnected to turn the instrument
off.
Measurement window
The main (numeric) measurement window continuously displays:
• Sensor measured values
• Measured sensor trends (for the last 10 minute to last hour)
• Measured sensor data alarm limits and other events
• Temperature
Touch screen
The user interface on the front panel is a touch screen providing easy selection through menus. All
the measurement, configuration, calibration and standard service routines can be called by pressing
buttons and menu bars on screen.
The display can be configured to only show a sensor measurement, or to show a parameterized
graphic representation of the last measurements.
English
13
Summary of Contents for ORBISPHERE 410
Page 179: ...ESD 2000 m 6562 ft 2000 m 179...
Page 180: ...GFCI GFI PE 100 240 VAC 5 VDC PE PE 1 U 2 3 2 4 1 U 2 3 4 2 5 180...
Page 181: ...1 3 4 5 6 7 1 2 3 6 4 4 5 4 1 4 6 7 1 1 5 USB A 2 USB B 4 6 2 3 Ethernet 7 1 4 8 181...
Page 182: ...NEMA IP NEMA 4X IP66 1 Ethernet 2 25 mm 8 mm 3 4 1 4 7 2 5 8 3 6 5 6 10 30 VDC 8 BINDER 182...
Page 184: ...1 4 2 3 1 2 3 3 1 2 3 4 1 2 3 4 3 4 5 4 2 6 1 7 ORBISPHERE J8 P8 4 185 J7 J8 6 186 7 186 184...
Page 188: ...32605 ORBISPHERE TC LED 10 1 menu Up Main Close Help Enter 188...
Page 190: ...1 Configure View Conf Channel temperature Grid y z Auto scale Clean 2 190...
Page 191: ...1 Config instrument Measurement Continuous 1 EC EC ppm ppb 5 0 1 2 3 HOT 90 2 TC TC EC TC 191...
Page 196: ...2 OK ideal current 3 O2 O2 30 O2 O2 O2 O3 196...
Page 197: ...O3 O2 O2 O3 O2 O2 O3 O3 O3 TC 1 Modify Modify 2 OK ideal current 3 197...
Page 198: ...ppm Cancel Validation Enable Enable Service done RS485 PROFIBUS DP USB HTTP TCP IP 198...
Page 199: ...Hach 199...
Page 224: ...2000 6562 2000 224...
Page 225: ...GFCI GFI PE 100 240 5 1 U 2 3 4 225...
Page 226: ...1 U 2 3 4 5 1 3 4 5 6 7 1 2 3 6 4 4 5 4 1 4 6 7 226...
Page 227: ...1 1 5 USB A 2 4 USB B 6 2 3 Ethernet 7 1 4 8 NEMA IP NEMA 4X IP66 1 Ethernet 2 25 8 3 4 227...
Page 228: ...1 4 7 2 5 8 3 6 5 6 10 30 8 BINDER 2 BINDER 1 10 30 2 3 4 5 6 10 30 7 10 30 8 3 228...
Page 231: ...Ethernet RJ 45 ethernet ethernet 1 227 P3 4 6 7 J8 6 7 8 J7 9 J8 J7 16 J7 8 231...
Page 233: ...J4 1 J5 2 J6 3 40 ORBISPHERE 29089 ORBISPHERE 32605 ORBISPHERE 233...
Page 234: ...10 Menu Up Main Close Help CAP Enter Security 234...
Page 244: ...last calibration variation 3 ppm Cancel Validation Services Enable Enable Service done 244...
Page 356: ......
Page 357: ......