61
ModBUS Register Information
Setup/Analog Output 2
Type
40031
Unsigned
Integer
1
R/W
Selects output type (0 = Linear output;
1 = PID control)
Setup/Analog Output 2
Transfer Value
40032
Float
2
R/W
Sets the transfer value
Setup/Analog Output 2
Filter
40034
Unsigned
Integer
1
R/W
Sets the output filter value in seconds
(0 to 120 sec)
Setup/Analog Output 2
0mA - 4mA
Select
40035
Unsigned
Integer
1
R/W
Selects 0mA/4mA for min output
(0 = 0mA; 1 = 4mA)
Setup/Analog Output
2/Linear
Min Setting
40036
Float
2
R/W
Sets the min output value
Setup/Analog Output
2/Linear
Max Setting
40038
Float
2
R/W
Sets the max output value
Setup/Analog Output
2/PID
Mode
40040
Unsigned
Integer
1
R/W
Sets the PID mode (0 = auto; 1 = manual)
Setup/Analog Output
2/PID
Manual Set
40041
Float
2
R/W
Sets the PID manual output value
(0.0 to 100.0%)
Setup/Analog Output
2/PID
Setpoint
40043
Float
2
R/W
Sets the PID setpoint
Setup/Analog Output
2/PID
Phase
40045
Unsigned
Integer
1
R/W
Sets the PID phase
(0 = Direct; 1 = Reverse)
Setup/Analog Output
2/PID
Proportional
Band
40046
Float
2
R/W
Sets the PID proportional band
Setup/Analog Output
2/PID
Integral Time
40048
Unsigned
Integer
1
R/W
Sets the PID integral time (min)
Setup/Analog Output
2/PID
Derivative
Time
40049
Unsigned
Integer
1
R/W
Sets the PID derivative time (min)
Setup/Relay 1
Source
40050
Unsigned
Integer
1
R/W
Selects data source for this relay
(0 = None; 1 = Real Time Clock;
2 = Sensor; 4 = Calculation)
Setup/Relay 1
Sensor Select
40051
Unsigned
Integer
1
R/W
Selects Sensor source when
Source = Sensor
(0 = Sensor1; 1 = Sensor2)
Setup/Relay 1
Measurement
Select
40052
Unsigned
Integer
1
R/W
Selects measurement on the Sensor
(0 = Meas1 .. 3 = Meas4)
Setup/Relay 1
Type
40053
Unsigned
Integer
1
R/W
Selects the relay type
(0 = Alarm; 1 = Control;
2 = Status; 3 = Timer; 4 = Event)
Setup/Relay 1
Transfer
Setting
40054
Unsigned
Integer
1
R/W
Selects the transfer value for the relays
(0 = De-energized; 1 = Energized)
Setup/Relay 1/Alarm
High Alarm
40055
Float
2
R/W
Sets the high alarm setpoint
Setup/Relay 1/Alarm
Low Alarm
40057
Float
2
R/W
Sets the low alarm setpoint
Setup/Relay 1/Alarm
High
Deadband
40059
Float
2
R/W
Sets the high alarm deadband
Setup/Relay 1/Alarm
Low
Deadband
40061
Float
2
R/W
Sets the low alarm deadband
Setup/Relay 1/Alarm
On Delay
40063
Unsigned
Integer
1
R/W
Sets the on delay time
Setup/Relay 1/Alarm
Off Delay
40064
Unsigned
Integer
1
R/W
Sets the off delay time
Setup/Relay 1/Control
Setpoint
40065
Float
2
R/W
Sets the controller setpoint
Table A-8 Controller ModBUS Registers (continued)
Group Name
Tag Name
Register # Data Type
Length
R/W
Description