Operation
5.6 Zero point setting
Select
Menu level
Accept
1.
MAIN MENU
2.
SENSOR-SETUP
3.
SELECT SENSOR
(if there is more than one sensor)
4.
CALIBRATION
5.
OFFSET
6.
OUTPUT MODE information
7.
POUR
0 SANDARD INTO MEAS.CHAMBER
(Dry measuring chamber and use filter set for
zero point calibration LZV325)
8.
PRESS ENTER
WHEN CONSTANT
x.xxx TRBFNU
9.
CALIBRATE
(x.xxxTRBFNU)
(Set zero point)
10.
CAL. READY
OUTPUT
ACTIVE