14
Installation
Fig. 3-10
Switch panel installation
Remove the sensor plug prior to the installation of the controller in the cut-out in
the switch panel.
1.
Remove the terminal strip J5 from the socket and unscrew the cable from the
terminal strip (Fig. 3-11).
2.
Undo the nuts for the sensor sockets on the inside of the housing and remove
the sensor sockets and cables (10) from the housing.
3.
After fastening the controller in the switch panel, re-fit the sensor socket with
the nuts and re-insert the terminal strip J5 after fitting the cable.
1.
Controller
6.
Oval-head bolt M6 x 1.0 x 20 mm (4)
2.
Rubber seal for switch panel installation
7.
Hex nut M6 x 1.0 (4)
3.
Switch panel, max. 9.5 mm thick
8.
Washer (4)
4.
Clip (2x) for switch panel installation
9.
Oval-head bolt M6 x 1.0 x 150 mm (4)
5.
Clip for fastening the controller
10. For installation, remove the sensor plug if necessary, see
below