High level calibration
Symptom
Possible solution
Signal not in range - calibration
value very different from last
calibration value
Check your theoretical sample O
2
content and reference configuration
If gas bottle calibration, completely open the sample flow adjustment valve
to avoid gas overpressure, hence wrong O
2
concentration
Stability not reached - signal not
stable
Wait for stability and check reference gas or liquid flow rate
Presence of liquid on LDO membrane. Check presence of liquid in flow
meter and if so, dry by injecting 20 mL of ethanol (EtOH) then 5 minutes of
N
2
gas
Instrument storage when not used
General guidelines
When not in use, ensure the instrument is turned OFF by pressing the ON/OFF switch (
No. 2
in
on page 8) until the green indicator light is extinguished.
C A U T I O N
If the instrument is to be stored in an environment where the temperature is likely to be 0°C (32°F) or below, then
to avoid any damage to the instrument make sure there is no liquid inside. Do this by first running warm water
through the instrument and dry by flowing dry air or N
2
through it. Then turn the sample flow valve (
No. 8
in
on page 8) to the OFF position.
Short term storage
For short term storage (between measurements or up to a maximum of 6 hours), leave the sample in
the instrument by turning the sample flow valve (
No. 8
in
on page 8) to the OFF position.
Overnight or weekend storage
When storing the instrument overnight or over a weekend, run clean water through the instrument to
prevent passageways from becoming clogged and then turn the sample flow valve (
No. 8
in
on page 8) to the OFF position. Wipe down the outside of the instrument with a clean damp cloth.
Long term storage
For long term storage (more than 1 week), run warm water through the instrument followed by 20 mL
of ethanol (EtOH). Dry by flowing dry air or N
2
through the instrument, and then turn the sample flow
valve (
No. 8
in
on page 8) to the OFF position. Wipe down the outside of the instrument with
a clean damp cloth. It is recommended to fully charge the battery prior to any long term storage.
Note: If the instrument has been in storage for more than 4 weeks, remember to fully recharge the battery pack
before use.
English
23
Summary of Contents for Lange ORBISPHERE 3100
Page 247: ...TFT 72 x 54 mm 2 83 x 2 13 ins 1 x USB 5 VDC 1 x RS232 0 5 V 3100 247...
Page 248: ...10 60 C Hach Lange 60 C 10 45 C Hach Lange Hach Lange 248...
Page 249: ...ORBISPHERE 3100 5 760 3 4mm 6mm 1 4 Hach Lange 249...
Page 251: ...4 OFF 1 1 4 USB 2 ON OFF 5 3 RS232 6 2 1 2 PURGE ON OFF PURGE 5 251...
Page 255: ...1 Cancel 5 2 USB 6 RFID 3 7 4 6 4 RFID USB RFID USB USB 10 USB Import Export Main Menu 255...
Page 256: ...Measurement stopped Enter 0000 0 9999 253 254 Enter Supervisor1 5678 256...
Page 257: ...1 7 2 8 USB USB 3 9 4 10 5 11 6 12 5 5 5 4 257...
Page 258: ...254 254 ID0 262 Enter Enter Cancel Enter 258...
Page 259: ...Enter 1 2 12 24 3 4 1 5 760 259...
Page 265: ...1 2 3 4 Nb Reset service timer Last service date 2 1 2 2 2 2 DC 2 DC LED 265...
Page 267: ...1 2 No 1 2 No 2 3 4 5 6 2 7 60 150 200 mL min PURGE 5 2 N2 267...
Page 269: ...No 8 2 251 1 20 mL EtOH N2 No 8 2 251 4 269...
Page 359: ...1 x USB 5 VDC 1 x RS232 0 5 V 3100 359...
Page 360: ...10 60 C Hach Lange 60 C 10 45 C Hach Lange 360...
Page 361: ...Hach Lange ORBISPHERE 3100 5 760 3 4mm 6mm 6 35mm 361...
Page 362: ...Hach Lange 1 362 2 2 1 363 364 3 367 4 371 5 3100 365 6 366 7 366 8 366 1 1 2 2 3 OFF ON 362...
Page 363: ...4 1 1 4 USB 2 5 3 RS232 6 2 1 2 363...
Page 367: ...Enter 6 1 Cancel 5 Enter 2 USB 6 RFID 3 7 4 6 4 RFID USB RFID USB USB 367...
Page 368: ...10 USB Measurement stopped Enter 0000 0 9999 366 366 Enter 368...
Page 369: ...Supervisor1 1 5678 1 7 2 8 USB USB 3 9 4 10 5 11 6 12 5 5 5 369...
Page 370: ...4 y 366 366 ID0 374 Enter Enter Cancel Enter 370...
Page 379: ...8 365 4 DG33228 1 2 1 2 2 3 4 5 6 2 7 379...
Page 380: ...60 150 200 mL min PURGE 5 2 2 LDO Hach Lange N2 CO2 50 mbar 2 LDO 20 mL EtOH 2 5 380...
Page 382: ......
Page 383: ......