• For tank and in-line applications the recommended flow rate should be above 200mL/min and
up to a maximum of having the flow adjustment valve fully open.
Note: The flow rates indicated with arrows on the flow meter (as illustrated below) are approximately
150mL/min (lower arrow) and 200mL/min (upper arrow). The silver bead gives an indication of the flow rate.
PC software installation
The instrument is delivered with default user and measurement configuration tables. However, to
personalize the instrument by setting up your own tables you will need to use the PC software which
is included on the USB key supplied with the tool kit.
Install the PC software by inserting the USB key into your PC and running the setup program
(entitled
setup.exe
) from the directory
Orbisphere 3100\Installation Files\PC Software
on the USB
drive. Follow the on-screen instructions and the software will be installed on your hard disk in a new
directory:
C:\Program Files\Hach Lange\3100 PC Software\
.
Once the software has been installed, click on the desktop icon on the PC to launch the application.
Create new user table
From the application’s
File
menu, select
New
and then
User Table
.
A default user is created automatically, with an ID of “0”, a User Name of “Default”, and a level of
“User”. None of these fields can be changed. The password is set automatically to “1234”, but this
can be changed.
Use the
Add
option to add new users, and
Delete
to remove existing users.
Copy
and
Paste
can
also be used to add new users, and
Cut
can be used to delete existing users. Double click on a field
to edit the contents.
Create new measurement configuration table
From the application’s
File
menu, select
New
and then
O2 Instrument
to create a new
measurement configuration table.
A default entry is created automatically, with an ID of “0”, a Location Name and Product Name of
“Default”. None of these fields can be changed.
Use the same functionality as described for the user table, to add entries to the table.
Transfer files to the instrument
When the two tables have been populated, they can be transferred to the instrument using a USB
storage device (typically a USB key).
1.
From the PC copy the files to the USB storage device in a top-level directory of
3100
. The files
will typically be located in:
C:\Program Files\Hach Lange\3100 PC Software\
with file
extensions of
.cdm
(for measurement configurations) and
.ndu
(for user tables).
Note: It is important that the file name extensions (.cdm and .ndu) are not changed as they will not be
recognized by the instrument software. Similarly, the files must be located in a top-level directory of 3100.
2.
With the instrument switched ON, insert the USB storage device into the USB connection on the
left side of the instrument, and press the USB icon on the instrument front panel.
3.
The first screen is for exporting files from the instrument to a USB storage device, so press the
right arrow to move to the next screen.
10
English
Summary of Contents for Lange ORBISPHERE 3100
Page 247: ...TFT 72 x 54 mm 2 83 x 2 13 ins 1 x USB 5 VDC 1 x RS232 0 5 V 3100 247...
Page 248: ...10 60 C Hach Lange 60 C 10 45 C Hach Lange Hach Lange 248...
Page 249: ...ORBISPHERE 3100 5 760 3 4mm 6mm 1 4 Hach Lange 249...
Page 251: ...4 OFF 1 1 4 USB 2 ON OFF 5 3 RS232 6 2 1 2 PURGE ON OFF PURGE 5 251...
Page 255: ...1 Cancel 5 2 USB 6 RFID 3 7 4 6 4 RFID USB RFID USB USB 10 USB Import Export Main Menu 255...
Page 256: ...Measurement stopped Enter 0000 0 9999 253 254 Enter Supervisor1 5678 256...
Page 257: ...1 7 2 8 USB USB 3 9 4 10 5 11 6 12 5 5 5 4 257...
Page 258: ...254 254 ID0 262 Enter Enter Cancel Enter 258...
Page 259: ...Enter 1 2 12 24 3 4 1 5 760 259...
Page 265: ...1 2 3 4 Nb Reset service timer Last service date 2 1 2 2 2 2 DC 2 DC LED 265...
Page 267: ...1 2 No 1 2 No 2 3 4 5 6 2 7 60 150 200 mL min PURGE 5 2 N2 267...
Page 269: ...No 8 2 251 1 20 mL EtOH N2 No 8 2 251 4 269...
Page 359: ...1 x USB 5 VDC 1 x RS232 0 5 V 3100 359...
Page 360: ...10 60 C Hach Lange 60 C 10 45 C Hach Lange 360...
Page 361: ...Hach Lange ORBISPHERE 3100 5 760 3 4mm 6mm 6 35mm 361...
Page 362: ...Hach Lange 1 362 2 2 1 363 364 3 367 4 371 5 3100 365 6 366 7 366 8 366 1 1 2 2 3 OFF ON 362...
Page 363: ...4 1 1 4 USB 2 5 3 RS232 6 2 1 2 363...
Page 367: ...Enter 6 1 Cancel 5 Enter 2 USB 6 RFID 3 7 4 6 4 RFID USB RFID USB USB 367...
Page 368: ...10 USB Measurement stopped Enter 0000 0 9999 366 366 Enter 368...
Page 369: ...Supervisor1 1 5678 1 7 2 8 USB USB 3 9 4 10 5 11 6 12 5 5 5 369...
Page 370: ...4 y 366 366 ID0 374 Enter Enter Cancel Enter 370...
Page 379: ...8 365 4 DG33228 1 2 1 2 2 3 4 5 6 2 7 379...
Page 380: ...60 150 200 mL min PURGE 5 2 2 LDO Hach Lange N2 CO2 50 mbar 2 LDO 20 mL EtOH 2 5 380...
Page 382: ......
Page 383: ......