3.
Remove the holding screw from the flow chamber assembly.
4.
Holding the assembly between thumb and forefinger, gently ease it out of the instrument and
discard.
5.
Replace the assembly with the new one from the maintenance kit, Put the new assembly back in
place in the instrument and secure with the holding screw. A guide rod is in place to ensure
correct positioning.
6.
Replace the front guard and secure in place with the 2 screws underneath the instrument.
7.
After changing the spot, a sensor calibration must be performed before using the instrument for
measurement purposes.
Troubleshooting
Measurement
Symptom
Possible solution
Display appears to be frozen - no
active cursor
Check measurement mode (continuous or sample)
Display appears to be frozen out
of range
When “Out of Range”, measurement interval is 60 seconds
Degassing in flow meter
Adjust flow rate to between 150 and 200 mL/min
Verify that the filter (if using) is not blocked
Measurement spiking
Ensure all tubing is tightly and correctly connected to the sample inlet and
outlet connections
Remove any residual air bubbles from the sample line by turning the sample
inlet/outlet valve to the
PURGE
position for 5 seconds
Response time too long
Use tubes supplied with instrument (low O
2
permeable material)
Check flow rate and connections
Prior to package sampling, purge the piercer with N
2
gas
Perform the monthly cleaning procedure
Channel out
Change LDO spot
Contact Hach Lange service
Silver bead does not move freely
in the sample
Perform the monthly cleaning procedure
Zero calibration
Symptom
Possible solution
Signal not in range - calibration
value very different from last
calibration value
Open sample flow adjustment valve completely
Calibration gas has too much oxygen, use quality 50 N
2
or CO
2
bottle
Oxygen leaks between reference bottle and instrument. Can be checked by
increasing gas flow rate and checking partial pressure (mbar) decrease
Stability not reached - signal not
stable
Wait for stability and check N
2
gas flow rate
Presence of liquid on LDO membrane. Check presence of liquid in flow
meter and if so, dry by injecting 20 mL of EtOH then 5 minutes of N
2
gas
22
English
Summary of Contents for Lange ORBISPHERE 3100
Page 247: ...TFT 72 x 54 mm 2 83 x 2 13 ins 1 x USB 5 VDC 1 x RS232 0 5 V 3100 247...
Page 248: ...10 60 C Hach Lange 60 C 10 45 C Hach Lange Hach Lange 248...
Page 249: ...ORBISPHERE 3100 5 760 3 4mm 6mm 1 4 Hach Lange 249...
Page 251: ...4 OFF 1 1 4 USB 2 ON OFF 5 3 RS232 6 2 1 2 PURGE ON OFF PURGE 5 251...
Page 255: ...1 Cancel 5 2 USB 6 RFID 3 7 4 6 4 RFID USB RFID USB USB 10 USB Import Export Main Menu 255...
Page 256: ...Measurement stopped Enter 0000 0 9999 253 254 Enter Supervisor1 5678 256...
Page 257: ...1 7 2 8 USB USB 3 9 4 10 5 11 6 12 5 5 5 4 257...
Page 258: ...254 254 ID0 262 Enter Enter Cancel Enter 258...
Page 259: ...Enter 1 2 12 24 3 4 1 5 760 259...
Page 265: ...1 2 3 4 Nb Reset service timer Last service date 2 1 2 2 2 2 DC 2 DC LED 265...
Page 267: ...1 2 No 1 2 No 2 3 4 5 6 2 7 60 150 200 mL min PURGE 5 2 N2 267...
Page 269: ...No 8 2 251 1 20 mL EtOH N2 No 8 2 251 4 269...
Page 359: ...1 x USB 5 VDC 1 x RS232 0 5 V 3100 359...
Page 360: ...10 60 C Hach Lange 60 C 10 45 C Hach Lange 360...
Page 361: ...Hach Lange ORBISPHERE 3100 5 760 3 4mm 6mm 6 35mm 361...
Page 362: ...Hach Lange 1 362 2 2 1 363 364 3 367 4 371 5 3100 365 6 366 7 366 8 366 1 1 2 2 3 OFF ON 362...
Page 363: ...4 1 1 4 USB 2 5 3 RS232 6 2 1 2 363...
Page 367: ...Enter 6 1 Cancel 5 Enter 2 USB 6 RFID 3 7 4 6 4 RFID USB RFID USB USB 367...
Page 368: ...10 USB Measurement stopped Enter 0000 0 9999 366 366 Enter 368...
Page 369: ...Supervisor1 1 5678 1 7 2 8 USB USB 3 9 4 10 5 11 6 12 5 5 5 369...
Page 370: ...4 y 366 366 ID0 374 Enter Enter Cancel Enter 370...
Page 379: ...8 365 4 DG33228 1 2 1 2 2 3 4 5 6 2 7 379...
Page 380: ...60 150 200 mL min PURGE 5 2 2 LDO Hach Lange N2 CO2 50 mbar 2 LDO 20 mL EtOH 2 5 380...
Page 382: ......
Page 383: ......