Clean the colorimeter cell as follows:
1.
Press the
MENU
key, then press the down
ARROW
key until the
MAINT
menu appears. Press
ENTER
.
2.
Press the down
ARROW
key to scroll to the
CLEAN
option. Press
ENTER
.
3.
Refer to
and locate the colorimeter.
4.
Remove the rubber plug on the top of the colorimeter assembly.
5.
Remove the magnetic stir bar using a bent steel paper clip as a retrieval
tool. See
6.
Wait until
CLEAN
flashes on the status line of the display, then fill
the colorimeter with 19.2 N Sulfuric Acid Standard Solution
(Cat. No. 2038-32).
Note:
Lower normality sulfuric acid will NOT be sufficient for cleaning the colorimeter.
7.
Allow the sulfuric acid to stand in the colorimeter for 15 minutes.
8.
Insert a wooden or paper cotton-tipped swab* into the opening and move
it up and down with a gentle scrubbing action, cleaning the interior
surfaces of the colorimeter cell. Refer to
. Clean the stir bar with
the cotton swab.
9.
Reinstall the stir bar.
10.
Replace the rubber plug on the top of the colorimeter assembly and make
sure the cover is on the viewing port of the colorimeter.
11.
Press the
EXIT
key to immediately return to normal operation; the
instrument will return to normal operation automatically after 60 minutes.
* Do not use a plastic swab when cleaning the colorimeter with sulfuric acid, the acid will dissolve the plastic.
Summary of Contents for Lange CL17
Page 1: ...Catalog Number 54400 18 CL17 CHLORINE ANALYZER INSTRUMENT MANUAL 06 04 4ed...
Page 2: ...Hach Company 2001 2004 All rights reserved Printed in the U S A...
Page 10: ...Visit http www hach com...
Page 12: ...Visit http www hach com...
Page 32: ...Visit http www hach com...
Page 46: ...Visit http www hach com...
Page 54: ...Visit http www hach com...
Page 58: ...Visit http www hach com...
Page 62: ...Visit http www hach com...
Page 68: ...Visit http www hach com...