Page 22
Electrical Connections
54400_install.fm
Section 1
Hach recommends conduit for two reasons:
1.
It is generally required by most local electrical codes.
2.
Use of metal conduit can improve immunity to lightning surges and
AC power transients.
Additionally, electrical and instrumentation standards require a local means of
removing power from the product.
The instrument is supplied with a
power on/off switch that is located inside the instrument enclosure.
To remove relay power from the instrument, an external customer-
supplied 5A fused switch or a 5A breaker is required.
In hard-wired electrical applications, the power and safety ground service
drops for the instrument should be no longer than 6 meters (20 feet) unless
metal conduit is used to shield the AC power wiring. The wire should be 18 to
12 AWG.
In applications where power cords are allowed by local electrical codes and
power surges and transients are not a great concern, a sealing-type strain
relief and power cord with three 18 gauge conductors (including a safety
ground wire) can be used. See
for strain relief assembly. The length
of the power cord must not exceed 3 meters (10 feet).
Figure 9
Assembling the Optional Power Cord Hardware
Description
Catalog Number
115 V Power cord with strain relief
54488-00
230 V Power cord with strain relief
54489-00
Wall section of the
instrument enclosure.
Summary of Contents for Lange CL17
Page 1: ...Catalog Number 54400 18 CL17 CHLORINE ANALYZER INSTRUMENT MANUAL 06 04 4ed...
Page 2: ...Hach Company 2001 2004 All rights reserved Printed in the U S A...
Page 10: ...Visit http www hach com...
Page 12: ...Visit http www hach com...
Page 32: ...Visit http www hach com...
Page 46: ...Visit http www hach com...
Page 54: ...Visit http www hach com...
Page 58: ...Visit http www hach com...
Page 62: ...Visit http www hach com...
Page 68: ...Visit http www hach com...