![Hach Lange CL17 Instrument Manual Download Page 50](http://html2.mh-extra.com/html/hach/lange-cl17/lange-cl17_instrument-manual_3163199050.webp)
Page 48
Unscheduled Maintenance
54400_maintain.fm
Section 3
3.2
Unscheduled Maintenance
3.2.1
Fuse Replacement
The T, 2.5A, 250V fuse used in this instrument is used for both 115V and
230V operation.
DANGER
Remove power from the
instrument when removing or
installing a fuse.
DANGER
Couper l'alimentation électrique de l'appareil pour retirer ou installer un fusible.
PELIGRO
Apagar la electricidad del instrumento al quitar o instalar fusibles.
GEFAHR
Beim Entfernen oder Einsetzen der Sicherung muss die Stromzufuhr zum Gerät
unterbrochen werden.
PERIGO
Desconecte a energia do instrumento ao remover ou instalar um fusível.
Replace as follows:
1.
Make sure there is no power supplied to the instrument. the instrument
power switch does not remove power from the fuses, Power must be
disconnected remotely to remove power from the fuses.
2.
Remove the customer access cover.
3.
Locate the fuse holders (near the terminal strip in the customer wiring
compartment). See
DANGER
For continued protection against
fire hazard, replace the fuse
only with a fuse of the same type
and rating.
DANGER
Pour assurer la protection contre les risques d'incendies, remplacez les fusibles
uniquement par des fusibles du même type et pour la même intensité.
PELIGRO
Para una continua protección contra incendios, reemplace los fusibles
únicamente por los del tipo y capacidad recomendados.
GEFAHR
Zur Wahrung des kontinuierlichen Brandschutzes dürfen die Sicherungen nur
mit Sicherungen des gleichen Typs und mit gleichen Stromkennwerten
verwendet werden.
PERIGO
Para proteçao contínua contra fogo, troque os fusíveis somente por fusíveis do
tipo especificado para a força da corrente.
4.
Remove the two fuses (F1 and F2) and replace them with two new fuses
with the same specifications (T, 2.5A, 250V). See
.
5.
Reinstall the customer access cover and resupply power.
!
!
Summary of Contents for Lange CL17
Page 1: ...Catalog Number 54400 18 CL17 CHLORINE ANALYZER INSTRUMENT MANUAL 06 04 4ed...
Page 2: ...Hach Company 2001 2004 All rights reserved Printed in the U S A...
Page 10: ...Visit http www hach com...
Page 12: ...Visit http www hach com...
Page 32: ...Visit http www hach com...
Page 46: ...Visit http www hach com...
Page 54: ...Visit http www hach com...
Page 58: ...Visit http www hach com...
Page 62: ...Visit http www hach com...
Page 68: ...Visit http www hach com...