Section 2
4.
Prepare a chlorine standard solution with a value between 3 and 5 mg/L.
Determine the value of the standard to the nearest 0.01 mg/L.
5.
Remove the container of zero reference water and substitute the chlorine
standard solution. Allow the analyzer to run on the standard solution for
approximately 10 minutes.
6.
When the readings are stable, enter the
SETUP
menu.
7.
Press the
DOWN ARROW
key until
CAL STD
is displayed. Press
ENTER
.
The current measured value is displayed.
8.
Press
ENTER
and edit the value. Press
ENTER
again to accept the
value.The measured value will be forced to the entered value. Press the
EXIT
key three times to return to normal display mode.
9.
Remove the standard and restore sample flow to the analyzer. The
instrument is now calibrated.
2.7.5.2 Calibration by Comparison
Note:
Make sure sample
chlorine concentration is relatively
stable before taking a grab
sample for laboratory analysis.
Perform the laboratory analysis as
quickly as possible.
Calibration by comparison involves analyzing the sample stream with a
reliable, accurate laboratory method, such as a DPD spectrophotometric or
amperometric titration method then setting the analyzer to match. Accuracy
specifications over the full 0 to 5 mg/L range cannot be assured unless the
sample chlorine concentration used for calibration is greater than 3.0 mg/L.
Setting the calibration at a lower concentration may provide accuracy at that
point but can cause large errors in other portions of the span.
Perform a calibration by comparison as follows:
1.
When the sample chlorine concentration is relatively stable, obtain a grab
sample for analysis.
2.
Perform a laboratory analysis (DPD spectophotometric or amperometric
titration method) on the grab sample and record the value.
3.
Enter the
SETUP
menu by pressing the
MENU
key followed by the
UP
ARROW
key and
ENTER
.
4.
Press the
DOWN ARROW
key until
CAL STD
is displayed. Press
ENTER
.
The current measured value is displayed.
5.
Press
ENTER
and edit the value to match the value from the laboratory
analysis. Press
ENTER
again to accept the value. The measured value will
be forced to the entered value.
6.
Press the
EXIT
key three times to return to normal display mode.
Summary of Contents for Lange CL17
Page 1: ...Catalog Number 54400 18 CL17 CHLORINE ANALYZER INSTRUMENT MANUAL 06 04 4ed...
Page 2: ...Hach Company 2001 2004 All rights reserved Printed in the U S A...
Page 10: ...Visit http www hach com...
Page 12: ...Visit http www hach com...
Page 32: ...Visit http www hach com...
Page 46: ...Visit http www hach com...
Page 54: ...Visit http www hach com...
Page 58: ...Visit http www hach com...
Page 62: ...Visit http www hach com...
Page 68: ...Visit http www hach com...