Page 40
Instrument Menu Structure
54400_start.fm
Section 2
•
PRIME
—This feature causes the analyzer to make 39 concurrent
accelerated cycles to prime the reagent lines. Press
ENTER
to activate. At
the end of 39 cycles, the analyzer will return to normal operation. To
terminate the prime cycle before all 39 cycles are complete, press
EXIT
.
The analyzer will return to normal operation.
•
CLEAN
—This feature stops the analyzer in the last portion of the
measurement cycle. Sample and reagents are left in the sample cell for
cleaning purposes. The analyzer remains in the clean mode for 60
minutes. To terminate the cycle earlier, press the
EXIT
key. The analyzer
will return to normal operation.
•
HOLD OUTPUTS
—This feature allows the alarms to be locked out and
the recorder output to be held at its present level for maintenance
purposes. Activate this feature as follows:
a.
Press
ENTER
then the
UP ARROW
key to activate for 60 minutes. The
alarm LED will be flashing.
b.
To deactivate the feature, and return to normal operation, press the
MENU
key then the
DOWN ARROW
key until
HOLD OUTPUTS
is
displayed.
c.
Press
ENTER
.
d.
Use the
DOWN
ARROW
key to select
Off
and press
ENTER
again.
2.7.5 Calibration
The CL17 Chlorine Analyzer is factory calibrated. A built-in electronic curve is
preprogrammed into the instrument.
This instrument does not require
recalibration unless specified by your regulatory agency for compliance
reporting purposes.
If you are required to do a two-point calibration or if your sample stream
chlorine concentration is typically less than 0.5 mg/L, follow the instructions in
. If your sample stream chlorine concentration is typically
greater than 0.5 mg/L, you can perform a Calibration by Comparison as
detailed in
.
If the calibration you perform requires you to adjust the CAL ZERO by more
than 0.2 mg/L, a SYSTEM WARNING caused by a MARGINAL ZERO error
will appear. Similarly, adjusting the CAL STD by more than 10% (up or down)
will produce a MARGINAL GAIN ERROR.
If either of these errors occur, reset the instrument to the default calibration
(see
), check the standards, and repeat the calibration. If the error
messages persist, contact the Customer Service department.
Summary of Contents for Lange CL17
Page 1: ...Catalog Number 54400 18 CL17 CHLORINE ANALYZER INSTRUMENT MANUAL 06 04 4ed...
Page 2: ...Hach Company 2001 2004 All rights reserved Printed in the U S A...
Page 10: ...Visit http www hach com...
Page 12: ...Visit http www hach com...
Page 32: ...Visit http www hach com...
Page 46: ...Visit http www hach com...
Page 54: ...Visit http www hach com...
Page 58: ...Visit http www hach com...
Page 62: ...Visit http www hach com...
Page 68: ...Visit http www hach com...