![Hach K1100 Basic User Manual Download Page 22](http://html2.mh-extra.com/html/hach/k1100/k1100_basic-user-manual_3278935022.webp)
1.
From the
Main
menu, select
Calibration
followed by
Gas sensor
and then
Configuration
. Make
sure the parameters are set as follows:
Option
Description
Auto-calibration
Not available for this sensor.
Manual-calibration
Make sure that the
Auto-End
box is checked.
Hold during calibration
Make sure this box is checked.
Interferences enabled during
calibration
Make sure this is disabled by unchecking the box.
Zero calibration bottle
Make sure this is disabled by unchecking the box as it is not relevant
for this sensor.
2.
Exit from the configuration screen by pressing
OK
.
3.
Select
Calibration
and perform a manual zero calibration as described in
on page 23. For high range sensors perform an additional 100% humid air calibration as
described in
100% humid air calibration (high range sensors only)
7.4.2.2 Manual calibration
Manual calibrations can be made at any time by following these steps:
1.
Remove the sensor from the sample line.
2.
Rinse the sensor head with clean water.
3.
Wipe the sensor head with a clean soft tissue to remove any excess moisture.
4.
If using the supplied calibration device insert the sensor into the sensor holder on top of the
calibration device. If not using the calibration device insert the sensor into the flow chamber.
5.
Flow the calibration sample through the calibration device or flow chamber as applicable. If using
the calibration device fully open the valve on the pressure reducer to give a gas flow rate of
0.1 L/min. If you are not using the supplied calibration device with pressure reducer the maximum
allowable inlet pressure must be no more than 2 bar absolute.
6.
Configure the calibration as described in
on page 22.
7.
Start the calibration as described in
on page 23,
on page 23 or
on page 24 depending on the
preferred calibration method.
7.4.3 Calibration configuration
Note: This option can also be invoked by pressing the Modify button in either the Zero calibration or High level
adjustment calibration screens.
1.
From the
Main
menu, select
Calibration
followed by
Gas sensor
and then
Configuration
.
Option
Description
Auto-calibration
Not available for this sensor.
Manual-calibration
When
Auto-End
is enabled a manual calibration will complete automatically when the
parameters defined in
Stop parameters
are reached. Press on
Configure
to set the
manual calibration parameters. If the calibration fails, the previous calibration
parameters remain unchanged and a warning message displayed.
Hold during
calibration
If checked this keeps the last measured value and stops updating the outputs during
the calibration or verification process. This avoids sending invalid information to any
connected device. At the end of a calibration, this hold remains on for a further
10 minutes to allow the system to stabilize.
Interferences
enabled during
calibration
This option enables the Chlorine or Salt interference during calibration. This option
must be used if there is chlorine in the calibration solution and if the chlorine
interference correction is enabled during measurement.
22
English
Summary of Contents for K1100
Page 308: ...K1100 M1100 28 mm 143 50 x 49 mm 0 74 kg 5 65 x 1 93 1 63 T 0 7 kg 2 3 3 1 3 2 308...
Page 309: ...3 3 ESD T 3 4 2000 m 6562 ft 2000 m 4 309...
Page 310: ...PE 100 240 VAC 5 VDC PE PE 4 1 1 U 2 3 2 4 4 2 1 U 2 3 4 2 5 310...
Page 311: ...4 3 1 3 4 5 6 7 1 2 3 6 4 4 5 4 1 4 6 7 311...
Page 317: ...4 8 4 8 1 3 5 10 15 20 15 2 Hach 4 8 2 1100 1100 150 mL min No 3 8 8 1 3 2 4 5 5 1 LED 317...
Page 318: ...10 1 5 2 5 3 menu Up Main Close Help 5 4 Enter 6 ID 319 1 OK OK 2 2 3 1007 ID 1234 OK 318...
Page 331: ...8 2 1 Ksv Ksv 328 1 2 3 8 2 2 1 2 3 8 2 3 331...
Page 332: ...1 2 3 4 332...
Page 386: ...3 3 3 4 2000 2000 4 386...
Page 387: ...PE 100 240 5 4 1 1 U 2 3 4 4 2 1 U 2 3 4 5 4 3 387...
Page 388: ...1 3 4 5 6 7 1 2 3 6 4 4 5 4 1 4 6 7 388...
Page 394: ...4 8 4 8 1 3 5 10 15 20 15 CO2 Hach 4 8 2 150 K1100 M1100 3 8 8 1 3 2 4 5 5 1 394...
Page 395: ...10 5 2 5 3 Menu Up Main Close Help 5 4 CAP Enter 6 Security 396 1 OK OK 2 2 395...
Page 409: ...8 2 6 CIP ClO2 8 2 1 Ksv Ksv 406 1 2 3 8 2 2 1 2 3 8 2 3 409...
Page 410: ...1 2 3 4 410...
Page 490: ...3 3 3 4 2000 m 6562 ft 2000 m 4 DC AC 490...
Page 491: ...100 240 V AC 5 V DC PE 4 1 1 U 2 3 2 4 4 2 1 U 2 3 4 2 5 491...
Page 492: ...4 3 1 3 4 5 6 7 1 2 3 4 4 2 5 4 1 4 6 7 492...
Page 498: ...4 8 4 8 1 3 5 10 15 20 15 CO2 Hach 4 8 2 K1100 M1100 150 mL min 3 8 8 1 3 2 4 5 5 1 LED 498...
Page 499: ...10 5 2 service 5 3 menu Up Main Close Help 5 4 CAP Enter 6 500 1 OK OK 2 2 499...
Page 513: ...ClO2 8 2 1 Ksv Ksv 510 1 2 3 8 2 2 1 2 3 8 2 3 513...
Page 514: ...1 2 3 4 514...
Page 540: ......
Page 541: ......