3.4 Remote Input Commands
Commands sent to the analyzer from a computer must be formatted as an ASCII character
string in uppercase or lowercase letters. Spaces are allowed, but not required. All
commands start with a three character string. If additional information is needed to set a
value, an equal sign “=” is entered, followed by the data desired. The command is
completed by sending a CR (OD Hex) [Example: To recall the alarm 1 setpoint:
A1S<CR>; to set alarm1 setpoint to 7.5: A1S=7.5<CR>]. Refer to Table 6 for a complete
command list.
All commands except cold start (CST) generate a response message indicating acceptance
or rejection of the command from the analyzer. If the command is rejected, the response
is “[?.”
Response messages from the instrument are preceded by a steering character, which the
computer program can test to determine if the line is a response message or if the line is a
routine printout. The steering character is an ASCII left bracket “[”(5B Hex).
Note: If the print interval is
set to zero (disabled), then
the only outputs generated
are response messages.
Programs (Basic, C, etc.) must be written for the computer to transmit the proper
command strings to the analyzer via the interface.
KEYBOARD LOCKED
ALARM 1 = 105.0 C LOW
ALARM 1 HYST = 0.5 C
ALARM 1 DELAY = 00 SECS
ALARM 2 = 16.00 PH HIGH
ALARM 2 HYST = 0.50 PH
ALARM 2 DELAY = 00 SECS
SIGNAL AVERAGE = 03 SECS
AUTO TEMPERATURE COMP
PRINT EVERY 00 MINS
REC MINIMUM = -2.00 PH
REC MAXIMUM = 16.00 PH
CONTROL SET PT = 7.00 PH
CONTROL PRESET = 50.0 %
CONTROL GAIN = 0.00
CONTROL RESET = 0.00 /HR
CONTROL RATE = 0.00 MIN
PH
MV
TEMP
ALARMS
6.35
69.7
49.0
6.34
70.5
48.9
*A1*
6.89
6.5
25.0M
*A1*
6.80
11.6
25.0M
*W5*7*11*
6.80
11.6
25.0M
*W7*8*
6.80
11.6
25.0M
6.81
11.3
5.0M
*ALO*
6.81
11.3
25.0M
*ALO*
6.35
69.4
49.0
*W8*
pH
OFFSET
CAL
7.00
27.8
25.0
*W6*11*
>16.00
<-2000.0
25.0
*A2*W9*11*
7.00
27.9
>105.0
*W11*
Summary of Contents for EC1000
Page 2: ...2...
Page 8: ...8...
Page 12: ...12...
Page 14: ...14...
Page 16: ...16...
Page 18: ...18 Figure 1 Controller Front Panel...
Page 20: ...20 Figure 2 Pump Module...
Page 24: ...24 Figure 6 pH Control Loop...
Page 54: ...54...
Page 56: ...56...
Page 58: ...58 Figure 12 Controller Installation Panel Mount...
Page 61: ...61 Figure 15 Installing the pH ORP Electrode Figure 16 Flow Thru Cell...
Page 70: ...70...
Page 73: ...73 6 4 Calibration See Section 2 3 Calibration Figure 21 Priming Tee Operation...
Page 74: ...74...
Page 80: ...80...
Page 83: ...83 Figure 24 Sensor Replacement...
Page 85: ...85 Figure 25 Hach One pH Process Electrode...
Page 89: ...89 Figure 26 Pump Module Plumbing Diagram...
Page 102: ...102...
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 108: ...108...
Page 112: ...112...
Page 116: ...116...
Page 122: ...122...
Page 123: ...123...
Page 124: ......