Table 18 Diagnostic Test Menu
Key
Option
Sub Menu
Description
(Press ENTER
and CLEAR
simultaneously
for Diagnostic
Menu)
DISP TST
NONE
Perform a display pattern test
RLY TEST
NONE
Perform a relay open/close test
SIO TEST
NONE
Perform a SIO loop-back test
REC TEST
ZERO
Drive recorder output to full scale
FS
Drive recorder output to zero
1/2 FS
Drive recorder output to 1/2 scale
CTRL TST
ZERO
Drive controller output to zero
FS
Drive controller output to full scale
1/2 FS
Drive controller output to 1/2 scale
D/A TEST
NONE
D/A - A/D loop test (jumper must be installed)
CAL DFT
PH DFT
Sets the pH probe calibration to default values
MV DFT
Sets the mV offset to the default of zero
TEMP DFT
Sets the temperature probe calibration to default values
COLD SRT
NONE
Perform a cold start
KEYLOCK
.NONE
Set keyboard lockout code
INP CAL
4MA MV
Calibrate 4 mA input current loop on mV input
20 MA MV
Calibrate 20 mA input current loop on mV input
4 MA TEM
Calibrate 4 mA input current loop on temperature input
20 MA TEMP
Calibrate 20 mA input current loop on temperature input
Summary of Contents for EC1000
Page 2: ...2...
Page 8: ...8...
Page 12: ...12...
Page 14: ...14...
Page 16: ...16...
Page 18: ...18 Figure 1 Controller Front Panel...
Page 20: ...20 Figure 2 Pump Module...
Page 24: ...24 Figure 6 pH Control Loop...
Page 54: ...54...
Page 56: ...56...
Page 58: ...58 Figure 12 Controller Installation Panel Mount...
Page 61: ...61 Figure 15 Installing the pH ORP Electrode Figure 16 Flow Thru Cell...
Page 70: ...70...
Page 73: ...73 6 4 Calibration See Section 2 3 Calibration Figure 21 Priming Tee Operation...
Page 74: ...74...
Page 80: ...80...
Page 83: ...83 Figure 24 Sensor Replacement...
Page 85: ...85 Figure 25 Hach One pH Process Electrode...
Page 89: ...89 Figure 26 Pump Module Plumbing Diagram...
Page 102: ...102...
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 108: ...108...
Page 112: ...112...
Page 116: ...116...
Page 122: ...122...
Page 123: ...123...
Page 124: ......