2.1.3
Restoring Default Settings
To clear all previous settings (not necessary on first time start-up), press
CLEAR
and
ENTER
simultaneously and the DISPlay TeST appears:
1. Press the
DOWN ARROW
key to scroll down through the display test menu to
COLD StaRT.
2. Press
ENTER
. This procedure returns all settings to the controller defaults. Table 2
shows all default values for the controller setup menu.
Table 2 Setup Menu Default Value
Menu
Options Submenu
Default
ALARM1
MODE
—
PH HI
SET PNT
—
16.00 PH
HYST
—
0.10 PH
DLY TIME
—
0 SEC
ALARM2
[same as ALARM1]
RECORDER
MODE
—
PH
WINDOW
MINIMUM
-2.00 PH
—
MAXIMUM
16.00 PH
OUT TRIM
0 TRIM
1=4.00
—
FS TRIM
1=20.00
CONTROL
MODE
—
PH
SET PNT
—
7.00 PH
DEAD BD
—
0.00 PH
PRESET
—
50.0%
GAIN
—
01.00
RESET
—
00.00/HR
RATE
—
00.00 MIN
TITR CUR
CUR
—
POINTS
03
OUT TRIM
0 TRIM
1=4.00
—
FS TRIM
1=20.00
INIT SET
°C COMP
TYPE
AUTO TC
—
SET PNT
25.0 °C
—
SOL TC
0.0 PH
SERIAL I/O
BAUD
1200
—
PARITY
NONE
—
DATA BIT
8
—
STOP BIT
2
—
PRT DLY
ENABLED
—
PRT INTR
00 MINS
OUTPUT 2
—
CONTROL
SIG AVG
—
1 SEC
Summary of Contents for EC1000
Page 2: ...2...
Page 8: ...8...
Page 12: ...12...
Page 14: ...14...
Page 16: ...16...
Page 18: ...18 Figure 1 Controller Front Panel...
Page 20: ...20 Figure 2 Pump Module...
Page 24: ...24 Figure 6 pH Control Loop...
Page 54: ...54...
Page 56: ...56...
Page 58: ...58 Figure 12 Controller Installation Panel Mount...
Page 61: ...61 Figure 15 Installing the pH ORP Electrode Figure 16 Flow Thru Cell...
Page 70: ...70...
Page 73: ...73 6 4 Calibration See Section 2 3 Calibration Figure 21 Priming Tee Operation...
Page 74: ...74...
Page 80: ...80...
Page 83: ...83 Figure 24 Sensor Replacement...
Page 85: ...85 Figure 25 Hach One pH Process Electrode...
Page 89: ...89 Figure 26 Pump Module Plumbing Diagram...
Page 102: ...102...
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 108: ...108...
Page 112: ...112...
Page 116: ...116...
Page 122: ...122...
Page 123: ...123...
Page 124: ......