Section 6
Reagents and Calibration Standards
6.1 Reagents
BioTector TOC analyzer uses the following reagents:
I.
Acid:
1.8 N Sulfuric Acid (H
2
SO
4
) Reagent containing 80 mg/l Manganese Sulfate Monohydrate
II.
Base:
1.2 N Sodium Hydroxide (NaOH) Reagent
Reagents should not contain high levels of organics, nitrates and phosphates. Ideally, the level of organics,
nitrates and phosphates should be less than 100
µ
g/l (ppb) in the deionized water used to prepare TOC
analyzer reagents.
Acid and Base reagents are stable up to 1 year. The reagents must be stored in a safe and secure location,
where the temperature does not drop below 2°C, in compliance with the local regulations. Table 6 below
summarizes the total days each BioTector TOC Analyzer reagent lasts at various system configurations:
Table 6 BioTector TOC Analyzer Reagent Consumption
REAGENTS
Container Size
(Liters)
TOTAL DAYS REAGENT LASTS
Acid
19
46
Base
19
46
Above table is derived from several online operation parameters such as 100% online time.
Recommended bunds (reagent spill trays) to contain above quantity reagents are 1x 50 Liters.
Summary of Contents for BioTector B3500e
Page 17: ...Page 17 Software Menu Diagram...
Page 43: ...Page 43 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 45: ...Page 45 Figure 6 BioTector oxygen concentrator layout...
Page 61: ...Page 61...
Page 73: ...Page 73 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 152: ...Page 152 Section 11 System Replacement and Spare Parts...
Page 161: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...