Section 10
Service and Maintenance
10.1 Weekly Maintenance
Check the air supply pressure:
A) An existing instrument air supply line
B) BioTector Compressor
Check and confirm that the air pressure is 1.5 bar on the filter pack pressure regulator.
Check the Sulfuric Acid level in the acid reagent container. Check the Acid Pump and confirm that
there are no leaks.
Check the Sodium Hydroxide (base) level in the base reagent container.
If installed check the sampling system:
Option 1: Confirm that the BioTector Vacuum Sampler or BioTector Venturi Sampler is
operating
Option 2: Confirm that the Sample Catch-Pot has sufficient sample flow, which guarantees a
fresh sample for each analysis cycle.
Option 3: Confirm that the Sample Pipe has sufficient sample flow, which guarantees a fresh
sample for each analysis cycle.
Check the Sample Pump and confirm that there are no leaks. Observe Sample Pump operation and
sample delivery from the sample point and confirm that there are no blockages inside sample lines.
Check all valves inside the BioTector and confirm that there are no leaks.
Check the filter mats in the Fan and Vent housings and confirm that they are not blocked.
Check the Sample Out and Exhaust lines outside the BioTector and confirm that there are no
damages and blockages.
Summary of Contents for BioTector B3500e
Page 17: ...Page 17 Software Menu Diagram...
Page 43: ...Page 43 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 45: ...Page 45 Figure 6 BioTector oxygen concentrator layout...
Page 61: ...Page 61...
Page 73: ...Page 73 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 152: ...Page 152 Section 11 System Replacement and Spare Parts...
Page 161: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...