100_ROTARY V STOP:ON
Rotary Valve stopped with its rotation
sensor on (sensor signal 1).
The sensor is faulty and it is always
showing on (sensor signal 1).
See Rotary Valve and Rotary Valve
Sensor in Section
Go to Simulate menu, set MFC
flow to 20 l/h and check the
rotation of the Rotary Valve.
Check ROTARY VALVE SENSOR
signals (1 “on” and 0 “off”) in the
Oxygen Controller Status menu as
the valve rotates.
Replace the valve.
When the warning is cleared, the
green LED, labeled as “Stepper”
on the Oxygen Controller Board,
should be on.
101_ROTARY V STOP:OFF Rotary Valve stopped with its rotation
sensor off (sensor signal 0).
The sensor is faulty and it is not
sensing the rotation of the pump.
Sensor in Section
Go to Simulate menu, set MFC
flow to 20 l/h and check the
rotation of the Rotary Valve.
Check ROTARY VALVE SENSOR
signals (1 “on” and 0 “off”) in the
Oxygen Controller Status menu as
the valve rotates.
Replace the valve.
When the warning is cleared, the
green LED, labeled as “Stepper”
on the Oxygen Controller Board,
should be on.
102_OZONE GEN FAULT
The current drawn through the Ozone
Generator has dropped below 0.5 Amp
for longer than 5 seconds. Fuse 4
(for
Cooler and Ozone Generator)
is blown on
the motherboard. See Section
details.
Check/replace the fuse 4 (F4) on
the motherboard.
Check fuse holder and confirm
that the fuse is snapped in tightly
into the fuse holder.
Press the “RESET” button on the
motherboard.
Power off the BioTector, wait 30
seconds and power up.
Go to Simulate menu, turn the
OZONE GENERATOR on and
adjust Ozone Current to 0.9 Amp
(±0.05 Amp) using the
potentiometer on the Ozone
Generator box.
103_OZONE FAN FAULT
If this optional device is fitted, the
voltage on the Ozone Generator Fan
has dropped below 1.5 Volts or has
exceeded 3.5 Volts for longer than 5
seconds.
Check the operation of the fan
inside the Ozone Generator.
Check wiring on terminals N01
and N03 on the Ozone Generator
Board.
Summary of Contents for BioTector B3500e
Page 17: ...Page 17 Software Menu Diagram...
Page 43: ...Page 43 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 45: ...Page 45 Figure 6 BioTector oxygen concentrator layout...
Page 61: ...Page 61...
Page 73: ...Page 73 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 152: ...Page 152 Section 11 System Replacement and Spare Parts...
Page 161: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...