8.3.5.2
Programmable Outputs
PWR BRD OUT 1-6
POWER BOARD OUTPUT1-6 menus typically contain output
settings related to external devices. These relay outputs, located
on the Power and Input/Output Board, are system optional
features.
POWER BOARD OUTPUT1-6 relays can be programmed to a
single output function or to multiple output functions as
described for OUTPUT 1-6 above. The programmed output
function or functions are marked with an asterisk “*” sign in these
menus as in the example for POWER BOARD OUTPUT 2,
which is programmed for STOP, FAULT, WARNING and NOTE
signals below.
In the example below, POWER BOARD OUTPUT 2 is activated
when STOP or FAULT or WARNING or NOTE signal is
triggered.
POWER BOARD OUTPUT 1 is a standard system relay
programmed to STOP, FAULT, WARNING and NOTE output
functions by default.
In BioTector configuration data download (or in All Data
download), the programmed output functions for each output is
tabulated and marked with asterisk “*” signs for clarity as
illustrated below.
PWR BRD OUT 2
DEFAULT STATE
N/E
STOP
*
FAULT
*
WARNING
*
NOTE
*
SAMPLER FILL
………………………
..
………………………
..
Summary of Contents for BioTector B3500e
Page 17: ...Page 17 Software Menu Diagram...
Page 43: ...Page 43 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 45: ...Page 45 Figure 6 BioTector oxygen concentrator layout...
Page 61: ...Page 61...
Page 73: ...Page 73 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 152: ...Page 152 Section 11 System Replacement and Spare Parts...
Page 161: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...