8.3.2 System Program
SYSTEM PROGRAM 1
SYSTEM PROGRAM 1 TIC + TOC e
SYSTEM PROGRAM 1 TIC + TOC defines the system
parameters for Oxidation Phase operation in TIC & TOC
analysis mode. In TIC & TOC analysis mode, the TIC and TOC
contents of a sample are measured consecutively by means of a
single sample injection into a single reactor.
Sample Pump
SAMPLE PUMP is a shortcut to Commissioning, Sample Pump
menu. See Section
for details.
ANALYZER ZERO 15s, 20 l/h
CO
2
analyzer establishes a zero offset level for the ANALYZER
ZERO time (15s by default) with the oxygen flow (20 l/h)
settings.
A “12_HIGH CO2 IN O2” fault is generated during this phase
and BioTector stops if the CO
2
analyzer’s reading is higher than
CO2 ZERO LINE plus the CO2 ZERO ALARM value (250ppm
by default) for three consecutive reactions. See Section
for details.
TIC ACID 20p
TIC ACID specifies the amount of acid reagent (in pulses) to be
injected into the reactor at range 1.
SAMPLE IN 7p, P-V
SAMPLE IN defines the amount of sample (in pulses) to be
injected into the reactor at operation range 1. P-V stands for
“Pinch Valve”, which identifies the type of valve used during
sample injection.
A single pulse pump operation corresponds to ½ revolution of
the pump roller.
Note that any change carried out on the factory SAMPLE IN
setting requires recalibration of the system.
SAMPLE BYPASS PURGE 10s, 7l/h
SAMPLE BYPASS PYRGE is the time sample lines are purged
with Oxygen gas flow.
SAMPLE INJECTION 7s, 5l/h
SAMPLE INJECTION is the time sample is pushed into the
Mixer Reactor using Oxygen gas flow.
SAMPLE LINE REV FILL 15p
SAMPLE LINE REV FILL is the Sample Pump reverse
operation, to fill the sample lines with oxidized liquid.
REACTOR VALVE FILL 3p
REACTOR VALVE FILL is the Sample Pump reverse operation,
to fill the Reactor Valve with oxidized liquid. This is carried out to
protect the Reactor Valve during oxidation process.
TIC SPARGE TIME 20s, 20 l/h
BioTector sparges and measures the TIC content of the sample
for the TIC SPARGE TIME (20s by default) with the oxygen flow
(20 l/h) settings.
If the TIC level does not drop below the TIC CHECK level
(10ppm CO
2
by default in Reaction Check menu) by the end of
TIC SPARGE TIME, BioTector automatically extends this time
(extension time) until the TIC level drops below 10ppm level.
A “50_TIC OVERFLOW” warning is generated and BioTector
keeps running if the TIC level does not drop below 10ppm by the
end of the maximum extension time, which is 300 seconds. The
maximum extension time is not a programmable function. It is a
fixed figure within the system software.
BASE 230p
BASE defines the amount of base (in pulses) to be injected into
the reactor for the reaction at range 1.
BASE OXIDATION 135s, 10 l/h
BioTector oxidizes the sample for the BASE OXIDATION time
(135s by default) with the oxygen flow (10 l/h) settings. If any
CO
2
is released during this phase, it is measured and added to
the TOC result because the default setting of PEAK
INTEGRATION is TOC+BASE OXID in Result Integration menu.
TOC ACID 210p
TOC ACID states the amount of acid (in pulses) to be injected
into the reactor for the TOC SPARGE phase at range 1.
Summary of Contents for BioTector B3500e
Page 17: ...Page 17 Software Menu Diagram...
Page 43: ...Page 43 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 45: ...Page 45 Figure 6 BioTector oxygen concentrator layout...
Page 61: ...Page 61...
Page 73: ...Page 73 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 152: ...Page 152 Section 11 System Replacement and Spare Parts...
Page 161: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...