This allows the connection of equipment in an intrinsically safe manner, even when the equipment is
not tested together. No additional certification testing is needed if each piece of equipment is
approved with entity parameters and all safety requirements are met.
All equipment connected to the instrument must have entity parameters that meet the requirements
shown on the control drawings. Be sure to include the parameters of all cables and connections.
The information on the control drawings gives directions for a safe installation. Installation reference
drawings that show suggested connections for specific installations may be provided upon request.
In no way are any such drawings to supersede the approved control drawings of the instrument and
any connected equipment. Obey all safety requirements of the control drawing before the
requirements of additional installation reference drawings.
Approved installation drawings
D A N G E R
Explosion hazard. Any installation or configuration that is not specifically shown on the control
drawings is not allowed. In all cases, the local authority with jurisdiction will make the final installation
decision.
Figure 2
is the approved control drawing. The certified drawing shows the ONLY approved method or
methods for installation for the instrument.
English
7
Summary of Contents for 1735 TRO
Page 2: ...English 3 Fran ais 10 Espa ol 17 24 31 38 2...
Page 8: ...Figure 2 Approved installation control drawing 8 English...
Page 15: ...Figure 2 Sch ma de contr le de l installation approuv Fran ais 15...
Page 22: ...Figura 2 Plano de control de instalaci n aprobado 22 Espa ol...
Page 24: ...1735 TRO ATEX EC Type EN 60079 0 EN 60079 11 IP64 IP20 I 3 24...
Page 25: ...ESD 2005 8 12 2002 96 EC 1735 TRO ATEX 1 1 1735 TRO 25...
Page 26: ...IIC 1 2 29 9 mm 15 mm 1735 TRO 1 26...
Page 27: ...1 1 2 2 29 27...
Page 28: ...2 28...
Page 29: ...2 29...
Page 30: ...IP64 IP20 1735 TRO 30...
Page 31: ...1735 TRO ATEX EC Type EN 60079 0 EN 60079 11 IP64 IP20 I 3 Web 31...
Page 32: ...ESD 2005 8 12 EU 2002 96 EC 1735 TRO ATEX Zone 1 Zone 1 1735 TRO 32...
Page 33: ...Zone 1 Group IIC 2 36 9 mm 15 mm 1735 TRO 1 33...
Page 34: ...1 1 2 2 36 34...
Page 35: ...2 35...
Page 36: ...2 36...
Page 37: ...IP64 IP20 1735 TRO 37...
Page 38: ...1735 TRO ATEX EC EN 60079 0 EN 60079 11 IP64 IP20 I 3 38...
Page 39: ...ESD 2005 8 12 EU 2002 96 EC 1735 TRO ATEX Zone 1 Zone 1 1735 TRO 39...
Page 40: ...Zone 1 Group IIC 2 43 9mm 15mm 1735 TRO 1 40...
Page 41: ...1 1 2 2 43 41...
Page 42: ...2 42...
Page 43: ...2 43...
Page 44: ...IP64 IP20 1735 TRO 44...
Page 45: ......