General 

Read the instruction manual with its security advices and warnings 

prior to the initial use

 of 

the  units,  to  secure  a  safe  and  conventional  use.  Keep  this  document  for  future  reference 
and pass it with the unit in case of resale. 
  
 

Grey shaded text marks features exclusively relevant for the fitness version.  
 

2

 

Classification of the unit

 

 

The units of the cardio line med are medical products according to guideline 93/42 EEC, categorized in 
class IIa. In the control range of the brake, the tolerance is less than 5% or with resistances 50 watts and 
below less than 3 watts.  

 

The items of the cardio line are consistent with the standard DIN EN ISO 20957-1 and DIN EN ISO 20957-5 
usage class S (commercial use/ gyms) of the accuracy class A +/- 10 % and designed to suit the needs of 
fitness gyms and preventive fitness. 

 

The items of the cardio line are consistent with the standard DIN EN ISO 20957 and designed for supervised 
range.  

 

Only  one  person  at  a  time  can  use  the  unit;  the  maximum  usage  weight  has  to  be  less  than  150  kg.  Units 
with the optional boosted version can be burdened up to 200 kg, see type label and additional instructions.

 

3

 

Safety warnings  

3.1

 

Safe assembly of the unit 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Fig.1: View of machine with indication of relevant elements 

 
 
 
 
 

Adjustment feet for 

levelling 

 

Pedals 

 

Handle bar

 

Monitor for individual 
settings and visualization of  
the training data 
 

Position of Type 

Label

 

/H-p-cosmos.local/dfs/erp$/article/cos30021ef-med600/manual/EN/20170926_cos30033ef-med600-ef-600_torqualizer_stair_instructions_for_use_EN.pdf

Summary of Contents for motion stair 600

Page 1: ...00 Version 3 0 Page 1 of 19 User manual for the motion stair 600 med and the motion stair 600 H p cosmos local dfs erp article cos30021ef med600 manual EN 20170926_cos30033ef med600 ef 600_torqualizer...

Page 2: ...12 Programs of the cardio line 10 4 13 Pulse controlled programs 15 4 14 PC Interface 15 5 Maintenance 16 5 1 Instructions for the error condition detection failure 16 5 2 Instructions for preventive...

Page 3: ...tandard DIN EN ISO 20957 1 and DIN EN ISO 20957 5 usage class S commercial use gyms of the accuracy class A 10 and designed to suit the needs of fitness gyms and preventive fitness The items of the ca...

Page 4: ...e not sure ask a supervisor Before the start of the workout it must be assured that no one is close to moving parts to avoid danger for a third person Above all hold unattended children off the units...

Page 5: ...y can be equipped with a Polar hand pulse system combined with a coded wireless signal transmitter With this technology two advanced measuring systems work together The hand pulse measurement has prio...

Page 6: ...a if the workout has been started using a storage media The Scan button changes the display mode parameters or graphical displays within the Dot Matrix on the monitor The Smart Key buttons function as...

Page 7: ...alories burned or the remaining calories to be burned are shown Level Display of the actual intensity 1 21 Storage media The symbol for the storage media appears whenever a member card is inserted int...

Page 8: ...the handling of the unit The storage media system supports the user in all aspects of training activities The system consists of a monitor unit with the storage media reader The slot used to enter th...

Page 9: ...r four seconds each After that the personal workout that was previously saved on the storage media is being started automatically and the workout session can begin The storage media can be removed aft...

Page 10: ...x x x x 5 Isokinetik 6 Therapy x x x X 7 Interval x x 8 K cal x x x x 9 Hill x x x 10 Random x 11 Distance x x x Tab 1 Programs of the cardio line med separated by units usage and biofeedback program...

Page 11: ...le By doing so the user can perform a so called cross training whereby the resistance is adjusted to the expected wattage during the workout and the ergometer will simply make fine tunings to the resi...

Page 12: ...cates the hill profile remaining and the actual position during the workout The intensity will be shown permanently as level or as wattage speed The heart frequency is shown if the trainee is using on...

Page 13: ...this symbol you start and end the programming of an interval sequence Stepless load increase through the time of the sequence Increasing by steps Finish of the interval programming symbol request of s...

Page 14: ...doesn t get used by an individual In this case you can ignore the results The display shows writing Program for some seconds The program is now stored on the media The storage media can now be used A...

Page 15: ...e as well as intensity alongside the recovery progress As a general rule intensity changes occur by increasing the workload e g increasing of the wattage At the therapy training the intensity gets var...

Page 16: ...for damages and tightness In general Clean the devices with mild purifier and a wet cloth to get off aggressive sweat Dry the device afterwards The usage of spray bottles is not allowed it has shown...

Page 17: ...ed control According to the transposition of the medical device directive 93 42 EEC into national law At elliptical and stairclimbing machines cross and stair Metrological Controls are not required by...

Page 18: ...cedures are possible 1 The service is conducted on site by our service employees 2 The required spare part s are shipped 3 An exchange unit is arranged and sent to the user The defective parts must be...

Page 19: ...ness GmbH Co KG Trippstadter Str 68 67691 Hochspeyer Germany Tel 49 0 6305 71499 0 Fax 49 0 6305 71499 111 info emotion fitness de www emotion fitness de H p cosmos local dfs erp article cos30021ef me...