Distance / Distance training

 

The  distance  training  works  in  a  rpm-dependent  mode,  which  means  that  the  intensity  adjusts  when  the  rpm 
changes. For this program, the parameters distance and intensity level/gear (1-21) must be pre-selected. At any 
time  during  the  workout,  the  level  can  be  adjusted  with 

+/-

.  The  chosen  distance  will  countdown  and  the 

intensity is  determined  from  among  the  21  intensity levels.  The  workload is shown  in  km/h in  the  display.  The 
time that is needed to complete the distance can be used as an indicator of performance. 
After completing the program, a display of the results is initiated. 
 
 
Watt sensitive (not available for fitness units) 
Watt constant training programs are likely to be used in the field of training control and therapy, because the 
intensity  control  can  be  very  accurate.  As  long  as  there  are  no  orthopaedic  reasons  for  therapy,  the  rpm-
independent power control of ergometers has a lot of advantages. 
However, if problems in the extremities used exist, the rpm-independent power control can lead to overstress. A 
reduction of the rpm increases the pedal impact a lot and can lead to overstress or program abort. 
To  eliminate  these  negative  aspects,  excessive  torsion  can  be  avoided  in  the  watt  sensitive  program  by  an 
intelligent program control. During the program setting, a rpm limit gets selected. When the limit gets undercut 
the pedal resistance gets redeemed immediately.   
Also  the  overshoot  of  the  limited  heart  rate  gets  controlled.  When  it  gets  overshot  for  a  longer  time,  a  watt 
reduction of 10% every 30 seconds is the result (see motion pulse manager brochure). 
The target resistance will be reached via a ramp during the first 2 minutes. That way the trainee gets used to the 
power and an adequate warm-up is assured. 
The biofeedback function is in this program active and gives feedback according to the target rpm.  
 
 
 

Therapy (not available for fitness units)

 

Also  in  rehabilitation  the  total  intensity  of  the  training  gets  controlled  with  the  intensity  steering  factors 
workout time as well as intensity, alongside the recovery progress. As a general rule intensity changes occur by 
increasing  the  workload  (e.g.  increasing  of  the  wattage).  At  the  therapy  training  the  intensity  gets  varied 
several times within the program. After a two minute warm-up in ramp shape, 6 intervals of one minute each, 
with one minute active rest (50% of the impact intensity) per interval follow.  
The maximum interval impact in watt gets limited as well as the rpm limit below which the pedal power gets 
clearly decreased. Also the overshooting of the pre-entered heart rate gets controlled. When overshooting for a 
longer time a reduction of the wattage of 10% every 30 seconds follows (see motion pulse manager brochure).  
A Cool Down function over 3 minutes rounds up the training. Overall the program lasts 17 minutes. 
The biofeedback function is also active in this program and gives feedback according to the target rpm.  
 
 

4.13

 

Pulse controlled programs 

Concerning the working of pulse controlled programs such as pulse program, Kcal training, Watt sensitive and 
Therapy, please refer to the brochure “motion pulse manager”. 

 

4.14

 

PC-Interface 

All ergometers are equipped standardly with an RS232 interface. This interface is not equipped with a galvanic 
isolation  or  external  power  supply.  By  this  the  ergometer  can  be  controlled  via  a  PC,  an  ECG,  spiroergometry 
device, aso, if both devices are using the same interface protocol.  Please note, that with medical use, the rules 
for electric safety must be obeyed. Optionally, a RS232 interface  with galvanic isolation can be  ordered (when 
the machine is being build), in order to limit the risk resulting from any malfunction of the connected device. 
Please contact your dealer, when you have more questions about this issue. 
If the monitor shall be active for a longer period  – especially prior to the start of a training/test  – a stabilized, 
medically  approved,  9  Volt  power  supply  unit  has  to  be  connected.  This  is  not  included  in  the  standard  Rs232 
setting. 
After  connecting  the  power  supply,  the  display  is  being  activated  for  3  Minutes  and  then  goes  into  standby. 
With the next push of a button, it is ready instantly. 

/H-p-cosmos.local/dfs/erp$/article/cos30021ef-med600/manual/EN/20170926_cos30033ef-med600-ef-600_torqualizer_stair_instructions_for_use_EN.pdf

Summary of Contents for motion stair 600

Page 1: ...00 Version 3 0 Page 1 of 19 User manual for the motion stair 600 med and the motion stair 600 H p cosmos local dfs erp article cos30021ef med600 manual EN 20170926_cos30033ef med600 ef 600_torqualizer...

Page 2: ...12 Programs of the cardio line 10 4 13 Pulse controlled programs 15 4 14 PC Interface 15 5 Maintenance 16 5 1 Instructions for the error condition detection failure 16 5 2 Instructions for preventive...

Page 3: ...tandard DIN EN ISO 20957 1 and DIN EN ISO 20957 5 usage class S commercial use gyms of the accuracy class A 10 and designed to suit the needs of fitness gyms and preventive fitness The items of the ca...

Page 4: ...e not sure ask a supervisor Before the start of the workout it must be assured that no one is close to moving parts to avoid danger for a third person Above all hold unattended children off the units...

Page 5: ...y can be equipped with a Polar hand pulse system combined with a coded wireless signal transmitter With this technology two advanced measuring systems work together The hand pulse measurement has prio...

Page 6: ...a if the workout has been started using a storage media The Scan button changes the display mode parameters or graphical displays within the Dot Matrix on the monitor The Smart Key buttons function as...

Page 7: ...alories burned or the remaining calories to be burned are shown Level Display of the actual intensity 1 21 Storage media The symbol for the storage media appears whenever a member card is inserted int...

Page 8: ...the handling of the unit The storage media system supports the user in all aspects of training activities The system consists of a monitor unit with the storage media reader The slot used to enter th...

Page 9: ...r four seconds each After that the personal workout that was previously saved on the storage media is being started automatically and the workout session can begin The storage media can be removed aft...

Page 10: ...x x x x 5 Isokinetik 6 Therapy x x x X 7 Interval x x 8 K cal x x x x 9 Hill x x x 10 Random x 11 Distance x x x Tab 1 Programs of the cardio line med separated by units usage and biofeedback program...

Page 11: ...le By doing so the user can perform a so called cross training whereby the resistance is adjusted to the expected wattage during the workout and the ergometer will simply make fine tunings to the resi...

Page 12: ...cates the hill profile remaining and the actual position during the workout The intensity will be shown permanently as level or as wattage speed The heart frequency is shown if the trainee is using on...

Page 13: ...this symbol you start and end the programming of an interval sequence Stepless load increase through the time of the sequence Increasing by steps Finish of the interval programming symbol request of s...

Page 14: ...doesn t get used by an individual In this case you can ignore the results The display shows writing Program for some seconds The program is now stored on the media The storage media can now be used A...

Page 15: ...e as well as intensity alongside the recovery progress As a general rule intensity changes occur by increasing the workload e g increasing of the wattage At the therapy training the intensity gets var...

Page 16: ...for damages and tightness In general Clean the devices with mild purifier and a wet cloth to get off aggressive sweat Dry the device afterwards The usage of spray bottles is not allowed it has shown...

Page 17: ...ed control According to the transposition of the medical device directive 93 42 EEC into national law At elliptical and stairclimbing machines cross and stair Metrological Controls are not required by...

Page 18: ...cedures are possible 1 The service is conducted on site by our service employees 2 The required spare part s are shipped 3 An exchange unit is arranged and sent to the user The defective parts must be...

Page 19: ...ness GmbH Co KG Trippstadter Str 68 67691 Hochspeyer Germany Tel 49 0 6305 71499 0 Fax 49 0 6305 71499 111 info emotion fitness de www emotion fitness de H p cosmos local dfs erp article cos30021ef me...

Reviews: