Clean the parts of the seat post and seat sleigh with silicone spray and wipe it dry afterwards. 

 

Check the pedal straps for damages. 

 

Only use original spare parts, therefore please contact us.

 

 
Warning:

 Retorque cranks and pedals 

After the first 10 hours of use the cranks and pedals need to get re-tightened.  
If they don´t get retightened, damage of the axle, pedals or cranks is possible and the warranty doesn´t cover 
the damage.

 

5.3

 

Maintenance for authorised specialists  

According to the transposition of the medical device directive 93/42 EEC into national law, emotion fitness calls 
for a maintenance check with examination of all relevant component parts by an authorized employee/partner. 
This maintenance check needs to be conducted 

every 12 months

 on all medical devices of emotion fitness. For all 

necessitated control steps please also consider your countries local requirements.

 

5.4

 

Performance of the metrological control pursuant Medical Devices Operator Ordinance  (not relevant for 
units of the fitness line) 

Medical  products  have  to  undergo  a  technical  measurement  check  every 

24  months

.  The  date  of  the  next 

control is shown on the control badge, which you will find next to the type plate. 
The  possible  calibration  in  the  course  of  the  technical  measurement  check  can  only  get  performed  by  an 
authorised company or professional. Contact your dealer in case you need a reference. 

5.5

 

Conduction  of  a  safety-related  control  (According  to  the  transposition  of  the  medical  device  directive 
93/42 EEC into national law) 

At elliptical- and stairclimbing machines (cross and stair) Metrological Controls are not required by the 
legislature as the users body weight does have a significant influence on the resulting power. On request 
emotion fitness offers to calibrate the brake output. 
 

The conduction of the Maintenance, Safety-Related Controls and Metrological Controls is in the sole 
responsibility of the operator! 
 

§ 2 subparagraph 2: This law also applies to the use, operation and maintaining of products that have not been 
placed on the market as a medical device but are being used with the intended purpose of a medical device as 
described in the national laws resulting from medical device directive 93/42EEC. 

 
They are regarded as medical devices in terms of the medical device guideline.

 

 

6

 

Technical data

 

 

The information about measurements and weight can vary because of technical changes and optional 
features 
measurements 

Weight 

Max. user weight 

105 cm x 60 cm x 145 cm (L/B/H) 

65 kg 

150 kg; boosted: 200 kg

 

 

System of resistance:  self-powered brake without the external power supply 

 

buffer battery:  3 x AA/ R03/ UM4 – mignon cell 

 

range of performance medical: 4-27 m/min sinking speed 

 

range of performance fitness: 4-27 m/min sinking speed 

 

freewheel: existing 

 

Following EU-guidelines are fulfilled: 
DIN EN ISO 20957-1 S-A-I 
DIN EN ISO 20957-5 S-A-I 
DIN EN 60601-1:2013 
2001/95/EG general product safety guideline  
93/42/EWG medical guideline (solely for medical units)

/H-p-cosmos.local/dfs/erp$/article/cos30021ef-med600/manual/EN/20170926_cos30033ef-med600-ef-600_torqualizer_stair_instructions_for_use_EN.pdf

Summary of Contents for motion stair 600

Page 1: ...00 Version 3 0 Page 1 of 19 User manual for the motion stair 600 med and the motion stair 600 H p cosmos local dfs erp article cos30021ef med600 manual EN 20170926_cos30033ef med600 ef 600_torqualizer...

Page 2: ...12 Programs of the cardio line 10 4 13 Pulse controlled programs 15 4 14 PC Interface 15 5 Maintenance 16 5 1 Instructions for the error condition detection failure 16 5 2 Instructions for preventive...

Page 3: ...tandard DIN EN ISO 20957 1 and DIN EN ISO 20957 5 usage class S commercial use gyms of the accuracy class A 10 and designed to suit the needs of fitness gyms and preventive fitness The items of the ca...

Page 4: ...e not sure ask a supervisor Before the start of the workout it must be assured that no one is close to moving parts to avoid danger for a third person Above all hold unattended children off the units...

Page 5: ...y can be equipped with a Polar hand pulse system combined with a coded wireless signal transmitter With this technology two advanced measuring systems work together The hand pulse measurement has prio...

Page 6: ...a if the workout has been started using a storage media The Scan button changes the display mode parameters or graphical displays within the Dot Matrix on the monitor The Smart Key buttons function as...

Page 7: ...alories burned or the remaining calories to be burned are shown Level Display of the actual intensity 1 21 Storage media The symbol for the storage media appears whenever a member card is inserted int...

Page 8: ...the handling of the unit The storage media system supports the user in all aspects of training activities The system consists of a monitor unit with the storage media reader The slot used to enter th...

Page 9: ...r four seconds each After that the personal workout that was previously saved on the storage media is being started automatically and the workout session can begin The storage media can be removed aft...

Page 10: ...x x x x 5 Isokinetik 6 Therapy x x x X 7 Interval x x 8 K cal x x x x 9 Hill x x x 10 Random x 11 Distance x x x Tab 1 Programs of the cardio line med separated by units usage and biofeedback program...

Page 11: ...le By doing so the user can perform a so called cross training whereby the resistance is adjusted to the expected wattage during the workout and the ergometer will simply make fine tunings to the resi...

Page 12: ...cates the hill profile remaining and the actual position during the workout The intensity will be shown permanently as level or as wattage speed The heart frequency is shown if the trainee is using on...

Page 13: ...this symbol you start and end the programming of an interval sequence Stepless load increase through the time of the sequence Increasing by steps Finish of the interval programming symbol request of s...

Page 14: ...doesn t get used by an individual In this case you can ignore the results The display shows writing Program for some seconds The program is now stored on the media The storage media can now be used A...

Page 15: ...e as well as intensity alongside the recovery progress As a general rule intensity changes occur by increasing the workload e g increasing of the wattage At the therapy training the intensity gets var...

Page 16: ...for damages and tightness In general Clean the devices with mild purifier and a wet cloth to get off aggressive sweat Dry the device afterwards The usage of spray bottles is not allowed it has shown...

Page 17: ...ed control According to the transposition of the medical device directive 93 42 EEC into national law At elliptical and stairclimbing machines cross and stair Metrological Controls are not required by...

Page 18: ...cedures are possible 1 The service is conducted on site by our service employees 2 The required spare part s are shipped 3 An exchange unit is arranged and sent to the user The defective parts must be...

Page 19: ...ness GmbH Co KG Trippstadter Str 68 67691 Hochspeyer Germany Tel 49 0 6305 71499 0 Fax 49 0 6305 71499 111 info emotion fitness de www emotion fitness de H p cosmos local dfs erp article cos30021ef me...

Reviews: