PSW Series User Manual
90
V-I Mode
Selects High Speed Priority or Slew Rate
Priority for CV or CC mode. The voltage or
current slew rate can only be edited if CC/CV
Slew Rate Priority is selected. The ISR indicator
will be lit for CC Slew Rate Priority and the
VSR indicator will be lit for CV Slew Rate
Priority.
Note: CC and CV Slew Rate Priority mode are
disabled when voltage/current output is set to
external control.
ISR
CC Slew Rate priority
VSR
CV Slew Rate priority
F-03
0 = CV high speed priority
1 = CC high speed priority
2 = CV slew rate priority
3 = CC slew rate priority
Rising Voltage
Slew Rate
Sets the rising voltage slew rate. Only
applicable if V-I Mode is set to CV Slew Rate
Priority.
F-04
0.01V/s~60V/s (PSW 30-XX)
0.1V/s~160V/s (PSW 80-XX)
0.1V/s~320V/s (PSW 160-XX)
Falling Voltage
Slew Rate
Sets the falling voltage slew rate. Only
applicable if V-I Mode is set to CV Slew Rate
Priority.
F-05
0.01V/s~60V/s (PSW 30-XX)
0.1V/s~160V/s (PSW 80-XX)
0.1V/s~320V/s (PSW 160-XX)