OPERATION
65
Bleeder Control
The Master unit is used to control the
bleeder settings. The bleeder resistors in all
the slave units are always turned off when
in parallel mode.
Output Voltage/
Output Current
Model
Single unit 2 units
3 units
PSW 30-36
30V
30V
30V
36A
72A
108A
PSW 80-13.5 80V
80V
80V
13.5A
27A
40.5A
PSW 160-7.2 160V
160V
160V
7.2A
14.4A
21.6A
PSW 30-72
30V
30V
30V
72A
144A
216A
PSW 80-27
80V
80V
80V
27A
54A
81A
PSW 160-14.4 160V
160V
160V
14.4A
28.8A
43.2A
PSW 30-108 30V
30V
30V
108A
216A
324A
PSW 80-40.5 80V
80V
80V
40.5A
81A
121.5A
PSW 160-21.6 160V
160V
160V
21.6A
43.2A
64.8A
Master-Slave Parallel Connection
Master-Slave
Connector
The Analog Control Connector is used for both
serial and parallel connections. The way the
connector is configured determines the
behavior of the master and slave units. For the
complete connector pin assignment, see page
102.