6.6
Error Message List
APS-1102 Programmable AC/DC Power Source
6-49
6.6
Error Message List
USB error messages are listed in
Table 6-15
below
.
Table 6-15. USB Error Message List
No. Message
Cause
0
“No error”
No error.
100 “Command
error”
Command was recognized as invalid due to any cause.
(General command error)
101 “Invalid
character”
Command was recognized as invalid due to invalid characters
(such as “VOLT&”) in command string.
102 “Syntax
error”
Command was recognized as invalid due to reception of
undefined command or parameter.
200 “Execution
error”
Error occurred when executing the command due to any cause.
(General execution error)
222 “Data out of range”
Error occurred when executing the command due to out of range
parameter.
350 “Too Many Errors”
New errors cannot be retained due to error queue overflow.
1
“Invalid with output on”
Error occurred when executing the command because the output
relay was on.
2
“Invalid with output off”
Error occurred when executing the command because the output
relay was off.
3
“Invalid in this mode”
Error occurred when executing the command due to inappropriate
output mode.
4
“Under Busy State”
Error occurred during execution of command due to busy device.
5
“Out of Limiter”
Error occurred during execution of command because limiter
value was exceeded.
6
“Invalid in this frequency”
Error occurred when executing the command due to inappropriate
oscillation frequency setting.
7
“Out of Input Signal
frequency”
Error occurred when executing the command due to out of the
external synchronization frequency range.
8
“Under Warning State”
Error occurred when executing the command because device was
operating in protected mode.