6.3
Command Descriptions
APS-1102 Programmable AC/DC Power Source
6-32
STATus:WARNing:ENABle
This sets the warning event enable register. The setting ranges from 0 to 31999.
Syntax:
STATus:WARNing:ENABle <NR1>
Parameters:
0 to 31999
Command example:
STAT:WARN:ENAB 1
Query
example:
STAT:WARN:ENAB?
Response format:
<NR1>
Response example:
1
STATus:WARNing[:EVENt]?
This returns the warning event register value. The response ranges from 0 to 31999.
Syntax:
STATus:WARNing[:EVENt]?
Parameters:
None
Query example:
STAT:WARN?
Response format:
<NR1>
Response
example: 0
STATus:WARNing:NTRansition
STATus:WARNing:PTRansition
This sets the warning transition filter (WRTF). The setting ranges from 0 to 31999.
When the NTR filter is set to 1, WRER is set to 1 when WRCR changes from 1 to 0.
When the PTR filter is set to 1, WRER is set to 1 when WRCR changes from 0 to 1.
When the NTR and PTR filters are set to 1, WRER is set to 1 when WRCR changes.
When the NTR and PTR filters are set to 0, changes in WRCR are not propagated to WRER.
Syntax:
STATus:WARNing:NTRansition <NR1>
STATus:WARNing:PTRansition <NR1>
Parameters:
0 to 31999
Command example:
STAT:WARN:NTR 16384
STAT:WARN:PTR
0
Query
example:
STAT:WARN:NTR?
STAT:WARN:PTR?
Response format:
<NR1>
Response example:
16384
0