Calibration/verification of Classe III Scale
In order to maintain the medical approval the digital scale must be calibrated/
verified according to national regulatory requirements at an accredited testing
institute.
Software version and change counter are verified on the hand control’s dis
-
play.
In daily use it can be verified that the system is approved for medical use by
a Notified Body by confirming that ”Software version counter” reads 1 and the
”Event counter” on the display corresponds with the counter values on the
verification label.
Both values are 1 from the factory and until the first re-verifications.
Menu
Scale Info 1-2
Version
01.00
Next
Scale Info 2-2
Changes
001
Next
Exit
1. Press any key on the hand control to activate the display.
2. Select “Menu” using the function key located immediately below the display.
3. Next, select
until the ”Version” menu appears in the display.
Version (information screen 1 of 2 in scale menu):
• The number before the decimal counts the software versions and larger
modifications;
• The number after the decimal counts minor software modifications and bug
fixes.
4. Select ”next” to read ”Changes”.
Changes (information screen 2 of 2 in scale menu): change counter
• Counts changes in software parameters such as calibration and values for
G factor.
5. Press EXIT to get back to the main menu.