15
(For example, pure water that is in accordance with the European
Pharmacopoeia or the United States Pharmacopoeia).
d) The air used for drying must be filtered by HEPA.
e) Regularly repair and inspect the disinfector.
9.5 Drying
If your cleaning and disinfection process does not have an automatic
drying function, dry it after cleaning and disinfection.
Methods
1. Spread a clean white paper (white cloth) on the flat table, point the
product against the white paper (white cloth), and then dry the product
with filtered dry compressed air (maximum pressure 3 bar). Until no
liquid is sprayed onto the white paper (white cloth), the product drying is
completed.
2. It can also be dried directly in a medical drying cabinet (or oven). The
recommended drying temperature is 80°C~120°C and the time should be
15~40 minutes.
Notes:
a) The drying of product must be performed in a clean place.
b) The drying temperature should not exceed 138°C;
c) The equipment used should be inspected and maintained regularly.
9.6 Inspection and maintenance
In this chapter, we only check the appearance of the product. After
inspection, ensure that there is no problem.
1. Check the product. If there is still visible stain on the product after
cleaning/ disinfection, the entire cleaning/disinfection process must be
repeated.
2. Check the product. If it is obviously damaged, smashed, detached,
corroded, it must be scrapped and not allowed to continue to be used.
3.Check the product. If the accessory is found to be damaged, please
replace it before use. And the new accessory for replacement must be
cleaned, disinfected and dried.
4.If the number of times of the product reaches the specified number of
times, please replace it in time.
9.7 Packaging
Install the disinfected and dried product and quickly package it in a
medical sterilization bag (or special holder, sterile box).
1.The package used conforms to ISO 11607;
2.It can withstand high temperature of 138°C and has sufficient steam
permeability;
3.The packaging environment and related tools must be cleaned regularly
to ensure cleanliness and prevent the introduction of contaminants;
4.Avoid contact with parts of different metals when packaging.