1. During use, please do not contact the heating part of the Hot Melting and
Filling Instrument. Before use, remember to install the Thermal Protector
Cap to prevent users or patients from scalding.
6.1 Choose Gutta Percha Injecting Needle
Choose suitable Gutta Percha Injecting Needle (20ga, 23ga or 25ga)
according to the situation of patient. And tighten the Gutta Percha Injecting
Needle and handpiece (Note: not too tight). When using, the Gutta Percha
Injecting Needle can rotate to suitable angle within the range of 360°in
clockwise direction and counterclockwise direction. And you can also use
wrench to pre-bend the needle and adjust it to a better operation angle as
per your needs.
Warning:
1. When install the Injecting Needle, please be sure that the device is off and
the head part of the device is cooling down. (About 5 minutes after shutdown
of the Hot Melting and Filling Instrument, the head part of it can cool down
to the temperature that allows people to touch.)
2. The pre-bending angle of the Injecting Needle cannot exceed 90 °,and do
not bend in the size transitional parts of the needle.
6.2 Choose the gutta-percha
Choose suitable gutta-percha for the device (the diameter of the gutta-
percha is less than 3mm,and the length of the gutta-percha is less than
17mm) .Remove the injecting needle firstly, put the gutta-percha into the
hole at the end of the injecting needle , and than screw the injecting needle
clockwise outo the handle and tighten it.
Note:
1.only one gutta-percha stick can be placed at a time.
2.Failure to fully fit the gutta-percha into he heating chamber will result in
function failure of the device.
6.3 Power on and Setting
6.3.1 Power on
After powering on with long press on “ON/OFF” button, the standby
interface display content is:
①
The serial number of mode;
②
Preset
temperature;
③
Injecting speed;
④
Battery power.
6.3.2 Mode setting
The device provides three groups of customizable parameter modes, the
parameters in each group of modes can be customized and saved.
After entering the normal standby interface, short press the mode button to
switch modes (3 modes can be customized). Long press the mode button
to enter the mode content mode setting interface of the currently mode,
short press the “ON/OFF” button to adjust the parameters, short press
the mode button to switch the adjustment items, short press the injecting
button to save the parameters and returns to the normal standby interface.