or avoid the generation of bubbles. Place the needle at the bottom of the
root canal, press the injecting button to squeeze out the gutta precha, then
slowly withdraw the needle tip until the coronal hole.
Warning
:
The maximum temperature of the applied part surface is 97°C . After
heating, the glue injection needle should not stay in the oral cavity for more
than 5 seconds to prevent scalding the patient.
6.5 Replacement of gutta-percha
When the gutta-percha is finished , you will hear a “di” sound. After
releasing the injecting button, the device will automatically return to initial
position, wait for the screen shows “Insert Gutta-Percha”, and then you
can insert a new gutta-percha.
When inserting a new gutta-percha , make sure that the front end of the
Hot Melting and Filling Instrument has cooled to room temperature to
prevent burns , and re-select a suitable gutta-percha to insert according to
chapter 3.2
.
6.6 Cleaning , Disinfection , Sterilization and Maintenance
After operation, the remaining materials in the heating chamber must be
cleaned, and the relevant accessories must be cleaned, disinfected and
sterilized. For details, see Chapter 9.
7 Charging instruction
7.1 Use corresponding charging base for charging: Connect the power
adapter to the charging base, and connect to power supply. And then
correctly place the Hot Melting and Filling Instrument in the charging
base. When the Hot Melting and Filling Instrument is not placed on the
charging base, the indicator will flashes in amber and green alternately.
When the Hot Melting and Filling Instrument is placed on the charging
base, if the charging is being charged, the amber indicator will be on
constantly. When the battery is full, the amber indicator will be off and the
green indicator will be on constantly. Under normal situation, the charging
takes about 2.5h.
7.2 The battery used in this product has no memory and can be used at any
time or charged at any time.
7.3 Before first use of this device, please charge it at least for 3 hours.
Warning
:
1. If the device is not in use for a long time, please unplug the adapter to
isolate the circuits electrically from the supply mains.
2. Do not perform medical functions during charging.
3. Equipment must be connected to an appropriate power source when loss
of power source would result in an unacceptable risk.
8 Safety precautions