2
1.6.2 Type of protection against electric shock: Class
Ⅰ
1.6.3 Degree of protection against electric shock: B type applied part
1.6.4 Degree of protection against harmful ingress of water: Ordinary
device (IPX0). Foot switch is anti-drip device (IPX1).
1.6.5 The contact duration time of applied part:
Nozzle: Less than 10 minutes
1.6.6 The temperature of the surface of nozzle may reach 45°C if it is
used at maximum power.
1.7 Main technical specification
1.7.1 Power adapter input: 100-240V~ 50/60Hz 1.1A Model: USE48-
300130SPA3
1.7.2 Main unit input: 30V 1.3A
1.7.3 Main unit insurance: T1.6AL 250V
1.7.4 Air inlet pressure: 5.5bar ~ 7.5bar(0.55MPa ~ 0.75MPa)
1.7.5 Main unit weight: 2.0kg
1.7.6 Main unit size: length×width×height 270mm×170mm×90mm
1.8 Operation environment
1.8.1 Environment temperature: + 5°C ~ + 40°C
1.8.2 Relative humidity: 30% ~ 80%
1.8.3 Atmospheric pressure: 70kPa ~ 106kPa
1.9 Intended place of use
Professional healthcare facility environment.
2 Installation and Operation
2.1 Main unit and main accessories display
2.1.1 Front view of main unit
Fig. 1 Front view of main unit
2.1.2 Back view of main unit