15
The sterilization time is at least 4 minutes at a temperature of 132°C /
134°C and a pressure of 2.0 bar ~ 2.3 bars.
Allow a maximum sterilization time of 20 minutes at 134°C.
Verification of the fundamental suitability of the products for effective
steam sterilization was provided by a verified testing laboratory.
Precautions:
1) Only products that have been effectively cleaned and disinfected
are allowed to be sterilized;
2) Before using the sterilizer for sterilization, read the Instruction
Manual provided by the device manufacturer and follow the
instructions.
3) Do not use hot air sterilization and radiation sterilization as this
may result in damage to the product;
4) Please use the recommended sterilization procedures for
sterilization. It is not recommended to sterilize with other sterilization
procedures such as ethylene oxide, formaldehyde and low temperature
plasma sterilization. The manufacturer assumes no responsibility for
the procedures that have not been recommended.
If you use the sterilization procedures that have not been
recommended, please adhere to related effective standards and verify
the suitability and effectiveness.
*Fractionation pre-vacuum procedure: a procedure for steam
sterilization by repeating pre-vacuum, the procedure used here is
steam sterilized by three pre-vacuums.
5 Maintenance, storage and transportation
5.1 Maintenance
5.1.1 It is recommended to install air drying device (such as freezing
dryer) in advance before using air polishing device. And check drying
device every day to ensure it work normally. Otherwise it may be easy
to cause prophylaxis powder agglomerating. And the agglomerating
will cause the device or handpiece blocked.
5.1.2 Check sealing ring on the handpiece, tail line, powder tank,
water bottle regularly. If finding defects such as rupture, deformation
or falling off, please refer to the instructions and replace them in time,
the accessories are equipped with sealing rings of corresponding
specifications.
5.1.3 Check the air inlet interface filter once a month to confirm
whether there is impurity in the filter. If there is impurity, replace