The model AP-A is intended for use in the electromagnetic environment specified
below. The customer or the user of the model AP-A should assure that it is used in
such an environment.
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic
environment - guidance
Conducted RF
lEC 61000-4-6
Conducted RF
lEC 61000-4-6
Radiated RF
lEC 61000-4-3
3 V, 0.15 MHz – 80
MHz, 6 V in ISM
bands between 0.15
MHz – 80 MHz, 80
% AM at 1 kHz
3 V/m and 10 V/m,
80 MHz – 2,7 GHz,
80 % AM at 1 kHz
385 MHz, pulse
modulation 18 Hz,
27 V/m;
450 MHz, FM +/- 5
kHz deviation 1 kHz
sine, 28 V/m;
710 MHz, 745 MHz,
780 MHz, pulse
modulation 217
Hz, 9 V/m;
810 MHz, 870 MHz,
930 MHz, pulse
modulation 18 Hz,
28 V/m;
1,720 MHz, 1,845
MHz, 1,970 MHz,
pulse
modulation 217 Hz,
28 V/m;
2,450 MHz, pulse
modulation 217 Hz,
28 V/m;
5,240 MHz, 5,500
MHz, 5,785 MHz,
pulse modulation
217 Hz, 9 V/m
3 V, 0.15 MHz – 80
MHz, 6 V in ISM
bands between
0.15 MHz – 80
MHz, 80 % AM at
1 kHz
3 V/m and 10 V/
m, 80 MHz – 2,7
GHz, 80 % AM at
1 kHz
385 MHz, pulse
modulation 18 Hz,
27 V/m;
450 MHz, FM +/-
5 kHz deviation 1
kHz sine, 28 V/m;
710 MHz, 745
MHz, 780 MHz,
pulse modulation
217 Hz, 9 V/m;
810 MHz, 870
MHz, 930 MHz,
pulse
modulation 18 Hz,
28 V/m;
1,720 MHz, 1,845
MHz, 1,970 MHz,
pulse
modulation 217
Hz, 28 V/m;
2,450 MHz, pulse
modulation 217
Hz, 28 V/m;
5,240 MHz, 5,500
MHz, 5,785 MHz,
pulse modulation
217 Hz, 9 V/m
Portable and mobile RF
communications equipment
should be used no closer to
any part of the models AP-
A, including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d=1.2×P
1/2
d=2×P
1/2
d=1.2×P
1/2
80 MHz to 800
MHz
d=2.3×P
1/2
800 MHz to 2.7
GHz
where P is the maximum
output power rating of the
transmitter In watts (W)
according to the transmitter
manufacturer and d Is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,a should be less than
the compliance level in each
frequency range.b
Interference may occur In
the vicinity of equipment
marked with the following
symbol: