Section 2
TM9711B-A-00
6 November 2013
2-1
2.0
9711B SPECIFICATIONS
2.1
Uncertainty Specifications
The specifications apply from Full Rate Current down to 10 % of the full rated current. For example the
300 A Range can be used from 300 A down to 30 A or the 100 A Range can be used down to 10 A.
Ensure current cables used can handle the full rated current.
Resistance
( )
Full Rated
Current Range
(A)
1 Year Accuracy
1, 2
(%)
Temperature
Coefficient
(
/
0
C)
Power Dissipation
at Rated Current
(mW)
10,000
0.000 1
± 0.01
5
0.001
1,000
0.000 1
± 0.01
5
0.01
100
0.001
± 0.01
5
0.1
10
0.01
± 0.01
5
1
1
0.1
± 0.01
5
10
0.1
1
± 0.01
10
100
0.01
10
± 0.01
10
1000
0.001
100
± 0.05
30
10 W
0.000 333 333
300
± 0.1
30
30 W
Note 1:
Calibrated in air at 23 ± 1
o
C referred to the unit of resistance as maintained by a National Metrology Institute, and expressed as a
deviation from nominal with a total uncertainty and coverage factor of k = 2.
Note 2:
A traceable report of calibration stating the measured values and uncertainty is provided with each unit at test currents of 1/10 of
rated current and full rated current to a maximum of 100A and a minimum of 0.1 mA,
2.2
General Specifications
G
ENERAL
S
PECIFICATIONS
Environmental
Temperature
Humidity
Operating
5
o
C to 38
o
C
40
o
F to 100
o
F
<70 % RH non-condensing
Storage
-20
o
C to 55
o
C
-40
o
F to 131
o
F
<90 % RH non-condensing
Exterior Dimension (Rack
Mount)
222 mm (H)
482 mm (W)
183 mm (D)
13.6 kg
8.7" (H)
19" (W)
7.2" (D)
30
lbs