Section 4
TM9711B-A-00
6 November 2013
4-4
Table 4-1 : List of Recommended Test Currents or Voltages For Resistance Values
Ampere
Input
Range
Resistance
Recommended
Test Currents
Foot
Notes
Comments
300
0.333 333 m
30 Adc
100 Adc
300 Adc
4
1, 4
6623A-300 Range Extender required .
Calibrate in order (minimum current to maximum
current)
100
10 Adc
100 Adc
1,2
6623A Range Extender required
Calibrate in order (minimum current to maximum
current)
10
1.0 Adc
10 Adc
1
6623A Range Extender required
Calibrate in order (minimum current to maximum
current)
1
0.1 Adc
1 Adc
2
6623A Range Extender required
Calibrate in order (minimum current to maximum
current)
0.1
0.01 Adc
0.1 Adc
3
Calibrate in order (minimum current to maximum
current)
0.01
0.001 Adc
0.01 Adc
3
Only 6622A Bridge Required (No Range Extender
needed) for Remaining Tests. Calibrate in order
(minimum current to maximum current)
0.001
0.000 1 Adc
0.001 Adc
3
Only 6622A Bridge Required (No Range Extender
needed) for Remaining Tests. Calibrate in order
(minimum current to maximum current)
0.000 1
0.000 1 Adc
3
Only 6622A Bridge Required (No Range Extender
needed) for Remaining Tests.
0.000 01
0.000 1 Adc
3
Only 6622A Bridge Required (No Range Extender
needed) for Remaining Tests.
1 Requires 6623A Series Range Extender
2 Requires either 6623A Series Range Extender or 6623A-2 (With Built in Supply)
3 Can use any of the recommended Bridges
4 Optional Calibration Points can be used in place of Standard Point or added to standard points for calibraiton.