GTE Industrieelektronik GmbH
V3.2UK 2/19
KMG-2000-G
Page 7 of 26
Once measuring has been completed, the
force/time graph is analysed automatically and
the values that are relevant to the standard are
determined.
•
Display measured values
Following evaluation, the course of the force over
the measurement period (5secs) is initially dis-
played and the dynamic force Fd (peak force) dis-
played as a value.
Normal 5sec. range display
Button F1
: To switch between ‘normal display’
(5sec measuring range) and ‘zoom display’
of the dynamic range (1sec measuring range)
Zoom display over 1sec dynamic range
Button F2
: To switch to display of 4 measured values:
Max. dynamic force:
Fd in N
Static force:
Fs in N
End force after 5 seconds:
Fe in N
Duration of dyn. force range:
td in ms
Button F3
: Cancellation of display mode
Once the third measurement has been taken, the
totals of 3 individual measurements required for
the normal measurement have been made. If but-
ton F3 is now actuated, the mean values of the 3
individual measurements are displayed. The mean
values are identified by the addition of an “m”. But-
tons F1 and F2 do not have any function as the
graphic representation of the mean values makes
no sense.
If the mean values are displayed, the display
mode can be cancelled using F3.
Measurement
is calculated
Please wait
Fd :
369N
3/1
Fs:
82N
< F1
Fe:
17N
> F2
Td:
128ms
F3
Fdm : 237N
3/3
Fsm:
58N
< F1
Fem:
12N
> F2
Tdm:
86ms
F3
362N
362N