Advanced Options for the TEOAE Diagnostic Unit
The Advanced Options menu permits modification of the test stimuli
and measurement values for the Custom Protocols of the TE
Diagnostic unit. Changes to the protocol should be made only by
qualified personnel, usually the administrator. If you are not familiar
with the use of these variables, do not attempt to change the protocols.
Changes to any of these characteristics may yield test results that
differ from those obtained in other test modes.
The Corti instrument comes with pre-programmed settings for the TE
and TE Custom protocols. See Appendix E for the manufacturer
settings of these protocols. Test protocol changes are saved in the
non-volatile memory.
Instructions for Customizing a Test Protocol
To enter the TEOAE Menu:
1)
Press
CHANGE at the main menu (Display 1)
2)
Using the
CHANGE
buttons, select the TEOAE
protocol you want to customize (the "TE 64s" protocol is not
customizable)
3)
Press
SETUP at the Protocol menu (Display 2)
4)
At the Clock menu (Display 15)
hold down
the
CHANGE
key for 3 seconds until the Ready light (green LED) turns off
5)
At the New BT Device menu (Display 18)
hold down
the
CHANGE key for 3 seconds until the Ready light (green
LED) turns off
The Averaging Time menu (Display 36) should appear on the display.
You are in the TEOAE menu and will be able to scroll the available
protocol parameters with the
NEXT button and make changes by
using the
LEFT or RIGHT
arrow keys to CHANGE the
selection.
NOTE:
If you push the
DOWN arrow key without holding it for 3
seconds, you will scroll through date and time, etc., rather than
accessing the displays that allow you to make changes to the custom
protocols.