Deleting Results
The Corti holds data in non-volatile memory. The data stays in
the memory even after data is printed or downloaded to the
Corti Data Manager. Data can be deleted through a few
methods, depending on the Save mode.
Save L/R Mode:
A single test for the Left ear and a single test for the Right
ear are held in memory. Data is deleted when a new test for
the left or right ear is acquired.
Data can be deleted using the Clear function in the System
Menu (page 32).
NOTE:
Following printing or data transfer to the PC software,
all tests saved in memory are marked for deletion and will be
permanently deleted when a new test is started. It is not
necessary to manually clear the results.
Save 250 Mode:
Data is deleted when new Patient Names are uploaded from
the Data Manager to the Corti (a warning is provided that
data will be deleted).
Data can be deleted using the Clear function in the System
Menu (page 32).
Data can be deleted in the Corti from the Data Manager when
the device is connected to the Data Manager (cable from
OAE Screener to PC). When Names is selected, the window
allows data to be deleted via the Clear Instrument button.
Data printed using Auto Print will be deleted when a new test
is started.
Printing to a Thermal Printer
Printing to an optional thermal printer is by way of wireless
connection. First establish wireless pairing between the Corti
instrument and the printer by following the instructions in the section
Instrument Settings - Wireless Device Pairing on page 31.
NOTE:
See the printer operating manual for instructions on using the
printer.
Following instructions provided with your printer, be sure the printer
is on and ready for communication/printing. From the Corti
instrument Main Menu (Display 1), press the
UP button to enter the
device connection screen (Display 13). Press the
CONNECT
button.
Display 13