Status/Error Messages
Display Messages:
Attach Probe
No probe is detected at the start of a test.
Device not
Responding
The printer is not responding to queries from the instrument.
BT Device Not
Found
The paired wireless device cannot be detected. The device may be
turned off or too far away.
BT Error #xxx
There is an error condition with the wireless device. Check the
status.
BT Not
Configured
The Corti instrument is not paired with any wireless device.
Fit Error
Cannot Obtain P
For a DP test, the desired level (L1 or L2) cannot be obtained within
allowable limits. User should refit the probe and retry the test.
Fit Error
Too High
For a DP test, the level of the calibration tone is too high. User
should refit the probe and retry the test.
Fit Error
Too Low
For a DP test, the level of the calibration tone is too low. User
should refit the probe and retry the test.
Limit Error
Overflow error during the calculation of the DFTs for a DP test.
User should repeat the test.
Memory Almost
Full
Saved tests are within 5 tests of the maximum limit.
Memory Full!
The maximum saved test limit is reached. The user will need to clear
the memory before any additional tests can be performed.
Power Low!
The battery charge level is too low for operation. The user must
charge the battery before additional tests can be performed.
Printer Error
Indicates a problem with the printer. Check the printer status.
Printer Paper Out!
Indicates printer paper has run out.
Time/Date Error
The clock is checked during power on to ensure it has not lost time
and been reset. In the case of clock reset, this message is shown. The
user should set the correct date/time.
Due for Service
Daily reminder will appear when the calibration due date has passed.
Indicator LEDs (lights):
NOISE / Orange
The indicator labeled ‘NOISE’ provides a visual indication (AMBER)
that the noise level measured during the test exceeds a nominal
threshold.
Also used to indicate some error conditions and when the outcome of
test is REFER, NOISY, or NO SEAL.
TEST / Yellow
The indicator labeled ‘TEST’ provides a visual indication (YELLOW)
that the selected test is being performed. This indicator will remain on
steady state during the test function.
READY / Green
The indicator labeled ‘READY’ lets the user know that the instrument
is not currently performing a test function and that it is available to
perform a test function.
CHARGE / Blue
The indicator labeled ‘CHARGE’ provides a visual indication
(BLUE) of the battery recharging function and battery status. The rate